FDA Adverse Event Injury Summary report: N

OPTIMIZER SMART MINI

MDR report key: 18414311 · Received December 28, 2023

Report

Report Number
3012563838-2023-00020
Event Type
Injury
Date Received
December 28, 2023
Date of Event
November 30, 2023
Report Date
December 28, 2023
Manufacturer
IMPULSE DYNAMICS USA, INC.
Product Code
QFV
UDI-DI
00810003380098
PMA / PMN Number
P180036/S007
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE EXPLANTED DEVICE PASSED ELECTRONIC CONTROLS TESTING AND NO ANOMALIES WERE PRESENT DURING EVALUATION OTHER THAN A SET SCREW BEING REMOVED. IT IS SUSPECTED THAT THE SET SCREW DURING INITIAL IMPLANTATION WAS NOT PROPERLY TORQUED DOWN, WHICH WOULD EXPLAIN THE "ASYMETRIC_PULSES" AND IMPEDANCE ERRORS THE DEVICE EXHIBITED DURING PREVIOUS INTERROGATIONS. THUS, THE ISSUE IS ATTRIBUTABLE TO IMPROPER SURGICAL TECHNIQUE.

Description of Event or Problem · 0

ON (B)(6) 2023, A PATIENT REPORTED TO THE IMPULSE DYNAMICS HOTLINE THAT THEIR IMPLANTED OPTIMIZER SMART MINI DEVICE WOULD NOT CHARGE AFTER MULTIPLE ATTEMPTS. THE PATIENT REPORTED THAT THEIR CHARGER WOULD EMIT THREE BEEPS BEFORE DISPLAYING AN ERROR CODE. THE PATIENT'S !PG WAS RESET AND INTERROGATED BY A MEMBER OF IMPULSE DYNAMICS FIELD STAFF. THE INTERROGATION REVEALED BOTH !PG LEADS HAD IMPEDANCE GREATER THAN 2000 OHMS, AND THE CAUSE OF THE DEVICE ENTERING DOWN MODE AND UNABLE TO BE CHARGED WAS A "THERAPY ERROR: DOWN_ASYMETRIC_PULSES". UPON RESET, THE !PG WAS ABLE TO DELIVER CCM THERAPY APPROPRIATELY AND BE CHARGED DESPITE THE PRESENCE OF PREMATURE VENTRICULAR CONTRACTIONS. THE PATIENT WAS SENT HOME AND ADVISED TO CALL BACK IF THE ERROR PERSISTED. ON (B)(6) THE PATIENT REQUESTED A FOLLOW-UP TO BE SEEN FOR THEIR IPG TO AGAIN BE RESET. ON (B)(6) THE PATIENT CONTACTED THE IMPULSE DYNAMICS HOTLINE TO AGAIN REPORT AN ERROR CODE BEING DISPLAYED ON THEIR CHARGER ATTRIBUTABLE TO THE IPG REVERTING TO DOWN MODE. THE IPG WAS RESET AGAIN ON DECEMBER 4 AND WAS SHOWN TO BE DELIVERING THERAPY APPROPRIATELY DESPITE BOTH LEAD IMPEDANCE READINGS REMAINING GREATER THAN 2000 OHMS. HOWEVER, DUE TO THESE RECURRENT PROBLEMS, A REVISION SURGERY WAS SCHEDULED AND OCCURRED ON DECEMBER 14. THE PATIENT WAS REPORTEDLY DOING WELL POST-OP, AND THE EXPLANTED IPG WAS SENT TO IMPULSE DYNAMICS USA, INC. FOR FURTHER EXAMINATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1361844 OPTIMIZER SMART MINI IMPLANTABLE PULSE GENERATOR QFV IMPULSE DYNAMICS USA, INC. CCM X11 H4881 00810003380098

Patients

Seq Age Sex Outcome Treatment
1 Male Other