FDA Adverse Event Malfunction Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1841259 · Received September 15, 2010

Report

Report Number
2023826-2010-00911
Event Type
Malfunction
Date Received
September 15, 2010
Date of Event
August 17, 2010
Report Date
August 17, 2010
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
PHYSICIAN ASSISTANT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION (OTHER): A MULTIFUNCTIONAL TEAM INVESTIGATED COMPLAINTS REGARDING LENS TEARS ASSOCIATED WITH CQ CARTRIDGE: THE RESULTANT CORRECTIVE ACTIONS INCLUDED IMPROVEMENTS IN MANUFACTURING, RELEASE TESTING, AND EVALUATION OF TEST RESULTS. ADDITIONALLY, THE INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT AS THE SURGEON INSERTED THE CQ2015A THREE PIECE LENS AND A HAPTIC GOT CAUGHT AND TORE IN THE INJECTOR. THE LENS WAS REMOVED WITHOUT ANY PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY CQ2015A NA

Patients

Seq Age Sex Outcome Treatment
1 80 YR INJECTOR: MODEL MSI-PM, LOT NUMBER UNKNOWN| CQ CARTRIDGE-FP MODEL, LOT NUMBER UNKNOWN