FDA Adverse Event
Malfunction
Summary report: N
COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA
MDR report key: 1841259
·
Received September 15, 2010
Report
- Report Number
- 2023826-2010-00911
- Event Type
- Malfunction
- Date Received
- September 15, 2010
- Date of Event
- August 17, 2010
- Report Date
- August 17, 2010
- Manufacturer
- STAAR SURGICAL COMPANY
- Product Code
- HQL
- PMA / PMN Number
- P990013
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- PHYSICIAN ASSISTANT
Narratives
Additional Manufacturer Narrative · 1
(B)(4). CONCLUSION (OTHER): A MULTIFUNCTIONAL TEAM INVESTIGATED COMPLAINTS REGARDING LENS TEARS ASSOCIATED WITH CQ CARTRIDGE: THE RESULTANT CORRECTIVE ACTIONS INCLUDED IMPROVEMENTS IN MANUFACTURING, RELEASE TESTING, AND EVALUATION OF TEST RESULTS. ADDITIONALLY, THE INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT AS THE SURGEON INSERTED THE CQ2015A THREE PIECE LENS AND A HAPTIC GOT CAUGHT AND TORE IN THE INJECTOR. THE LENS WAS REMOVED WITHOUT ANY PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA | INTRAOCULAR LENS | HQL | STAAR SURGICAL COMPANY | CQ2015A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | INJECTOR: MODEL MSI-PM, LOT NUMBER UNKNOWN| CQ CARTRIDGE-FP MODEL, LOT NUMBER UNKNOWN |