- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION, RESULTS: (ARTERIAL AND VENOUS OCCLUSION); (NARROW AORTIC NECK). CONCLUSIONS: (NARROW AORTIC NECK).
Description of Event or Problem · 1
AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5 CM ABDOMINAL AORTIC ANEURYSM APPROX 2 MONTHS AGO. THE AORTIC NECK WAS 12-14 MM LONG AND 18-20 MM IN DIAMETER. THE VESSELS WERE SMALL. THE PT HAD A RIGHT ACCESSORY RENAL ARTERY THAT WAS COVERED WITH THE STENT GRAFT; HOWEVER, THERE WAS STILL PERFUSION. THE CT SCAN REPORT NOTED A RIGHT RENAL ARTERY INFARCTION AND NO ENDOLEAKS WERE REPORTED. NO INTERVENTION WAS PERFORMED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00418689 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Required Intervention |