OLYMPUS MECHANICAL LITHOTRIPTOR
Report
- Report Number
- 8010047-2010-00175
- Event Type
- Injury
- Date Received
- September 16, 2010
- Date of Event
- November 6, 2009
- Report Date
- August 18, 2010
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- LQC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
OLYMPUS MADE SEVERAL ATTEMPTS TO OBTAIN ADD'L INFO REGARDING THE REPORT FROM THE PHYSICIAN SAID TO HAVE PERFORMED THE PROCEDURE, BUT DID NOT RECEIVE ADD'L INFO REGARDING THE REPORTED EVENT. THE SPECIFIC MODEL, LOT NUMBER OF THE MECHANICAL LITHOTRIPTOR ALLEGEDLY INVOLVED IN THE REPORTED EVENT REMAINS UNK. NO DEVICES INVOLVED IN THIS REPORT HAS BEEN RETURNED TO OLYMPUS FOR EVAL. THE EXACT CAUSE OF THE PT'S OUTCOME COULD NOT BE CONCLUSIVELY DETERMINED. IF ADD'L AND SIGNIFICANT INFO BECOMES AVAILABLE LATER THIS REPORT WILL BE UPDATED. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.
THE USER FACILITY REPORTED THAT DURING AN ENDOSCOPIC RETROGRADE. CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE WITH A SPHINCTEROTOMY, THE USER UTILIZED AN OLYMPUS MECHANICAL LITHOTRIPTOR TO CRUSH A BILE DUCT STONE. THE USER WAS ABLE TO CAPTURE THE STONE IN THE BASKET, BUT WHILE ADVANCING THE METAL SHEATH, IT WAS REPORTED THAT THE USER DISCOVERED THAT THE METAL HANDLE OF THE MECHANICAL LITHOTRIPTOR WAS MALFUNCTIONING (THE METAL WHEEL OF THE HANDLE WOULD ONLY CLICK BUT NOT TIGHTEN); THE USER REMOVED THE LITHOTRIPTOR HANDLE, CUT THE PIPE AND REMOVED BOTH THE MECHANICAL LITHOTRIPTOR AND THE SCOPE FROM THE WIRE. THE USER UTILIZED A NON-OLYMPUS MECHANICAL LITHOTRIPTOR (SOEHENDRA LITHOTRIPTOR) AND FED THE WIRES THROUGH THE CABLE AND THEN ATTACHED THE SOEHENDRA CRANK. IT WAS REPORTED THAT THE USER HAD ATTACHED THE WIRES TO THE FAR END OF THE HANDLE SO THAT WHEN HE TURNED THE HANDLE OF THE SOEHENDRA DEVICE, IT BROKE THE WIRES AT THE LEVEL OF THE ROTATING BAR. THE PT WAS REPORTEDLY LEFT WITH WIRES EXTENDING FROM HIS MOUTH FOR SEVERAL DAYS AND WAS TRANSFERRED TO A SECOND UNIDENTIFIED FACILITY FOR FURTHER INTERVENTION. THE PT ALLEGEDLY SUSTAINED PERMANENT INJURY TO HIS MOUTH AND THROAT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS MECHANICAL LITHOTRIPTOR | LITHOTRIPTOR | LQC | OLYMPUS MEDICAL SYSTEMS CORPORATION | UNK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SOEHENDRA LITHOTRIPTOR| BASKET WIRES MODEL UNK |