FDA Adverse Event Malfunction Summary report: N

CE INFUSOR LV10, 12 PACK

MDR report key: 1840849 · Received September 22, 2010

Report

Report Number
6000001-2010-03430
Event Type
Malfunction
Date Received
September 22, 2010
Date of Event
August 11, 2010
Report Date
August 30, 2010
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K041738
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL EXAMINATION WAS PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A LEAK. THE ROOT CAUSE WAS DETERMINED TO BE A RUPTURED RESERVOIR. THE DEVICE IS A SINGLE USE DEVICE AND WAS DISCARDED. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA # IDC-CAPA-(B)(4). BATCH REVIEW DETERMINED THAT NO NONCONFORMANCE REPORTS WERE DOCUMENTED FOR THIS LOT DURING MANUFACTURING. TREND REVIEW DETERMINED THAT SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT AN INFUSOR LV 10 WAS LEAKING DURING FILLING. THE DEVICE WAS BEING FILLED WITH 230 MILLILITERS OF NORMAL SALINE WHEN THE LEAK WAS OBSERVED. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CE INFUSOR LV10, 12 PACK PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 09K035

Patients

Seq Age Sex Outcome Treatment
1