FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 1840833 · Received September 22, 2010

Report

Report Number
6000001-2010-03421
Event Type
Malfunction
Date Received
September 22, 2010
Date of Event
May 20, 2010
Report Date
June 23, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION DISCOVERED AND CONFIRMED THE REPORTED CONDITION OF A FAILURE CODE 806:10, AND THE ROOT CAUSE WAS DETERMINED TO BE A FAULTY PUMP HEAD MODULE ON CHANNEL A. THE CHANNEL A PUMP HEAD MODULE WAS REPLACED TO REPAIR THE REPORTED CONDITION. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE PUMP WHICH EXPERIENCED FAILURE CODE 806:10. IT WAS NOT SPECIFIED WHEN OR AT WHAT POINT IN THE PROCESS THE REPORTED CONDITION OCCURRED. REVIEW OF THE DEVICE EVENT HISTORY DETERMINED THAT THIS CONDITION INTERRUPTED DELIVERY. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 5.04.00, WHICH IS CLASSIFIED AS UNREMEDIATED. NO ADDITIONAL INFORMATION IS AVAILABLE.