FDA Adverse Event Malfunction Summary report: N

TERUMO CDI 500 BLOOD PARAMETER MONITOR

MDR report key: 1840726 · Received September 17, 2010

Report

Report Number
1828100-2010-01646
Event Type
Malfunction
Date Received
September 17, 2010
Date of Event
June 4, 2010
Report Date
September 17, 2010
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DRY
PMA / PMN Number
K972962
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION IN PROGRESS, BUT NOT YET CONCLUDED.

Description of Event or Problem · 1

DURING ROUTINE TESTING OF THE DEVICE AT THE REPAIR CENTER, THE USER REPORTED THE MONITOR FAILED TO POWER ON. THE SINGLE BOARD COMPUTER AND BATTERIES WERE REPLACED TO RESOLVE THE PROBLEM. THE MONITOR WAS ORIGINALLY RETURNED FOR FAILURE TO POWER ON. SINCE THE EVENT OCCURRED AT THE SERVICE CENTER, THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TERUMO CDI 500 BLOOD PARAMETER MONITOR ON LINE BLOOD GAS MONITOR DRY TERUMO CARDIOVASCULAR SYSTEMS CORP. 500AHCT

Patients

Seq Age Sex Outcome Treatment
1