FDA Adverse Event Malfunction Summary report: N

ZIMMER TRABECULAR METAL HUMERAL STEM

MDR report key: 1840694 · Received September 17, 2010

Report

Report Number
1822565-2010-00718
Event Type
Malfunction
Date Received
September 17, 2010
Date of Event
December 19, 2008
Report Date
January 15, 2008
Manufacturer
ZIMMER, INC.
Product Code
HSD
Removal / Correction Number
1822565-7/26/2010-006-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL SUMMARY: DURING TRABECULAR METAL HUMERAL AND THE TRABECULAR METAL REVERSE SHOULDER PROCEDURES, INSTRUMENTS PREPARE THE BONE FOR A TOLERANCE RANGE FROM SLIGHT PRESSFIT TO SLIGHT CLEARANCE. BECAUSE THESE STEMS HAVE PROXIMAL TRABECULAR METAL COMBINED WITH A PROXIMAL TAPER THEY ACHIEVE ADEQUATE STABILITY THROUGH PROXIMAL FIXATION ALONE. IT HAS BEEN DETERMINED THAT DISTAL PRESS IS NOT REQUIRED AND THAT DISTAL PRESS COULD LEAD TO DIFFICULTY SEATING THE IMPLANT DEPENDANT ON BONE QUALITY, HUMERAL GEOMETRY, AND SURGEON REAMING. WHILE A CAUSE FOR THIS SPECIFIC CLINICAL EVENT CANNOT BE ASCERTAINED FROM THE INFO PROVIDED, THE COMMON CLINICAL PRESENTATION AND THE DATE OF THE OCCURRENCE SUGGESTS THAT THE EVENT IS RELATED TO THE ISSUES FOR WHICH ZIMMER IS IMPLEMENTING CORRECTIVE ACTION. REFERENCE REMOVAL REPORT 1822565-7/26/2010-006-R FOR ADDITIONAL INFO REGARDING THE CORRECTIVE ACTION TAKEN. NO PRODUCT WAS RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS WAS ALSO NOT POSSIBLE AS THE PRODUCT AND/OR LOT NUMBERS REQUIRED FOR RETRIEVAL WERE UNAVAILABLE. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE AVAILABLE INFO, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFO BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC. CONSIDERS THE INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT IS REPORTED THAT THE STEM WOULD NOT FULLY SEAT IN THE CANAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ZIMMER TRABECULAR METAL HUMERAL STEM SHOULDER PROSTHESIS HSD ZIMMER, INC.

Patients

Seq Age Sex Outcome Treatment
1