HARMONIC ACE 23CM W ERG HANDLE
Report
- Report Number
- 3005075853-2010-05449
- Event Type
- Malfunction
- Date Received
- September 22, 2010
- Date of Event
- August 30, 2010
- Report Date
- September 1, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). THE ANALYSIS SHOWED THAT THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED. THE DEVICE WAS ACTIVATED WITH THE GENERATOR AND AN ERROR CODE 5 WAS DISPLAYED. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN AN ERROR CODE 5 OR BLADE 'LOCKOUT' LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE
IT WAS REPORTED THAT DURING AN OPEN PROCEDURE, THE BLADE WAS BROKEN OFF ON THE MAYO TABLE. SINCE THE BLADE WAS BROKEN OFF OUTSIDE THE PATIENT, NO PIECES WERE LEFT INSIDE THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION FROM AN IMPLANT FORM THAT THIS DEVICE, RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) LEADS WERE REPLACED DUE TO ELECTROGRAM NOISE AND LOSS OF CAPTURE. THE DEVICE WAS EXPLANTED AND BOTH LEADS WERE SURGICALLY CAPPED. A NEW SYSTEM WAS IMPLANTED IN THE RIGHT PECTORAL REGION. NO DEFIBRILLATION THRESHOLD TESTING WAS PERFORMED DURING THIS PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 23CM W ERG HANDLE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | G4T16T |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |