STERLING BALLOON DILATATION CATHETER
Report
- Report Number
- 2134265-2010-04157
- Event Type
- Malfunction
- Date Received
- September 21, 2010
- Date of Event
- August 27, 2010
- Report Date
- August 28, 2010
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- DQY
- PMA / PMN Number
- K053116
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4): IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4)
IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) TREATMENT PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN A CALCIFIED AND MODERATELY TORTUOUS SHUNT VEIN. THE PHYSICIAN ADVANCED THE 5MM X 40MM STERLING BALLOON CATHETER. ON THE FIRST INFLATION THE BALLOON WAS INFLATED TO 6ATMS AND RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STERLING BALLOON DILATATION CATHETER | CATHETER, PERCUTANEOUS | DQY | BOSTON SCIENTIFIC - MAPLE GROVE | H74939032504040 | 13015534 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INTRODUCER SHEATH: 4F MOSQUITO |