FDA Adverse Event Malfunction Summary report: N

STERLING BALLOON DILATATION CATHETER

MDR report key: 1840373 · Received September 21, 2010

Report

Report Number
2134265-2010-04157
Event Type
Malfunction
Date Received
September 21, 2010
Date of Event
August 27, 2010
Report Date
August 28, 2010
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
DQY
PMA / PMN Number
K053116
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) TREATMENT PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN A CALCIFIED AND MODERATELY TORTUOUS SHUNT VEIN. THE PHYSICIAN ADVANCED THE 5MM X 40MM STERLING BALLOON CATHETER. ON THE FIRST INFLATION THE BALLOON WAS INFLATED TO 6ATMS AND RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERLING BALLOON DILATATION CATHETER CATHETER, PERCUTANEOUS DQY BOSTON SCIENTIFIC - MAPLE GROVE H74939032504040 13015534

Patients

Seq Age Sex Outcome Treatment
1 INTRODUCER SHEATH: 4F MOSQUITO