FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II 24GAX0.75IN PRN SLM NPVC

MDR report key: 18401319 · Received December 27, 2023

Report

Report Number
3014704491-2023-00847
Event Type
Malfunction
Date Received
December 27, 2023
Date of Event
November 15, 2023
Report Date
March 25, 2024
Manufacturer
BD (SUZHOU)
Product Code
FOZ
UDI-DI
00382903830787
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

1. DHR/BHR REVIEW(LOT#3052741): 1)THIS BATCH OF PRODUCTS WERE ASSEMBLED AT INTIMA II AUTO LINE (B)(6) 2023, AND PACKAGED AT R240 PACKAGE LINE IN (B)(6) 2023. WORK ORDER QUANTITY WAS (B)(4) EA. 2)REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MEET THE PRODUCT SPECIFICATIONS. 3)REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 2. NO DEFECTIVE SAMPLE AND PHOTO HAVE BEEN RECEIVED FOR THE COMPLAINT. 3. 45PSI LEAKAGE TEST IS CARRIED OUT ON THE RETAINED SAMPLE OF THIS BATCH, AND NO LEAKAGE IS FOUND. PLEASE REFER TO ATTACHMENT FOR THE TEST REPORT. 4. NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. CONCLUSION(S): NO ABNORMALITY IS FOUND ON PROCESS AND RETAINED SAMPLE, AND NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. AS THE DAMAGE STATE OF THE COMPLAINED SAMPLE CANNOT BE IDENTIFIED, THE ROOT CAUSE OF THE LEAKAGE CANNOT BE DETERMINED. THE PLANT WILL CONTINUE TO PAY ATTENTION TO SUCH DEFECTS. H3 OTHER TEXT : SEE NARRATIVE.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II 24GAX0.75IN PRN SLM NPVC LEAKED THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ON (B)(6) 2023, WHILE ADMINISTERING AN INDWELLING NEEDLE PUNCTURE INFUSION TREATMENT TO PATIENT HAN SHIYIN, THE END OF THE INDWELLING NEEDLE LEAKED, PREVENTING PROPER FLUID INTAKE AND REQUIRING RESETTING OF THE INDWELLING NEEDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2265212 BD INTIMA-II 24GAX0.75IN PRN SLM NPVC INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 3052741 00382903830787

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown