FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 1840104 · Received September 15, 2010

Report

Report Number
1720753-2010-03080
Event Type
Malfunction
Date Received
September 15, 2010
Date of Event
September 1, 2010
Report Date
September 15, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE PS3 POWER SUPPLY WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE VERTICAL LIFT COLUMN ON THE 9800 SYSTEM WOULD INTERMITTENTLY NOT WORK DURING A CASE. THE 9800 SYSTEM HAD TO BE REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER SYSTEM. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1