FDA Adverse Event Malfunction Summary report: N

INFINITI VISION SYSTEM OZIL

MDR report key: 1840065 · Received September 15, 2010

Report

Report Number
2028159-2010-01747
Event Type
Malfunction
Date Received
September 15, 2010
Date of Event
August 16, 2010
Report Date
August 16, 2010
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K082845
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER ORDERED A NEW FOOTSWITCH. THE FOOTSWITCH AND CABLE HAVE BEEN RECEIVED AND IN-HOUSE TESTING IS IN PROGRESS. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY WHEN ADDITIONAL REPORTABLE INFO BECOMES AVAILABLE. (B)(4).

Description of Event or Problem · 1

ADVERSE EVENT(S): "NO PT INJURY REPORTED" (NO CONSEQUENCES OR IMPACT TO PT). PRODUCT PROBLEM(S): "SYSTEM MESSAGE DISPLAYED" (DEVICE DISPLAYS ERROR MESSAGE). THE NURSE REPORTED A SYSTEM MESSAGE DISPLAYED DURING SET UP. THE SYSTEM MESSAGE NOTED A FOOTSWITCH FAILURE. ADDITIONAL INFO RECEIVED FROM THE NURSE STATED ONE CASE WAS DELAYED ABOUT 45 MINUTES. THE FACILITY WAS USING ONE OPERATING ROOM, INSTEAD OF TWO FOR A SHORT TIME. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITI VISION SYSTEM OZIL PHACOFRAGMENTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CENTER INFINITI OZIL NA

Patients

Seq Age Sex Outcome Treatment
1 UNK