FDA Adverse Event Malfunction Summary report: N

SPINAL SET 25GA 3-1/2IN

MDR report key: 18400458 · Received December 27, 2023

Report

Report Number
3003152976-2023-00554
Event Type
Malfunction
Date Received
December 27, 2023
Date of Event
December 11, 2023
Report Date
May 8, 2024
Manufacturer
BECTON DICKINSON
Product Code
MIA
UDI-DI
00382904050849
PMA / PMN Number
K210978
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PHOTO RECEIVED BY OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, THE IMAGE SHOWS THE STYLET ON ONE SIDE, THE CANNULA, AND THE CANNULA HUB SEPARATED FROM THE CANNULA. THEREFORE, THE INCIDENT IS CONFIRMED. ADDITIONALLY, PHYSICAL SAMPLE WAS RECEIVED AND DECONTAMINATED FOR EVALUATION. THE PHYSICAL SAMPLE WAS EVALUATED USING MAGNIFICATION, NO DAMAGE OR DEFECTS WERE OBSERVED. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 2210027 ,NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE. NINE RETAINED SAMPLES FROM THE SAME LOT WERE EVALUATED, NO DEFECTS WERE OBSERVED ON THE NEEDLE OR ANY OTHER COMPONENTS OF THE DEVICE. PRODUCT UNDERGOES A SERIES OF TESTING AND INSPECTIONS THROUGHOUT THE MANUFACTURING PROCESS THE ENSURE THE QUALITY AND FUNCTIONALITY OF THE DEVICE, INCLUDING VERIFICATION THE NEEDLE IS FREE FROM DAMAGE OR DEFECTS AND ALL CRITICAL DIMENSIONS ARE WITHIN SPECIFICATION. ALL INSPECTIONS FOR LOT 2210027 WERE COMPLETED ACCORDING TO PROCEDURE, NO ANNOTATIONS WERE NOTED RELATED TO THE REPORTED INCIDENT. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO IDENTIFY A ROOT CAUSE RELATED TO OUR MANUFACTURING PROCESS AT THIS TIME.

Additional Manufacturer Narrative · 0

PR (B)(4): INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

CORRECTION ON IMDRF CODE: B15 - ANALYSIS OF DATA PROVIDED BY USER/THIRD PARTY.

Additional Manufacturer Narrative · 0

PHOTO RECEIVED BY OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, THE IMAGE SHOWS THE STYLET ON ONE SIDE, THE CANNULA AND THE CANNULA HUB SEPARATED FROM THE CANNULA. THEREFORE, THE INCIDENT IS CONFIRMED. HOWEVER, PHYSICAL SAMPLES ARE REQUIRED TO FURTHER ANALYZE AND EVALUATE THE PRODUCT TO DETERMINE A ROOT CAUSE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 2210027 , NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE. NINE RETAINED SAMPLES FROM THE SAME LOT WERE EVALUATED, NO DEFECTS WERE OBSERVED ON THE NEEDLE OR ANY OTHER COMPONENTS OF THE DEVICE. PRODUCT UNDERGOES A SERIES OF TESTING AND INSPECTIONS THROUGHOUT THE MANUFACTURING PROCESS THE ENSURE THE QUALITY AND FUNCTIONALITY OF THE DEVICE, INCLUDING VERIFICATION THE NEEDLE IS FREE FROM DAMAGE OR DEFECTS AND ALL CRITICAL DIMENSIONS ARE WITHIN SPECIFICATION. ALL INSPECTIONS FOR LOT 2210027 WERE COMPLETED ACCORDING TO PROCEDURE, NO ANNOTATIONS WERE NOTED RELATED TO THE REPORTED INCIDENT. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO IDENTIFY A ROOT CAUSE RELATED TO OUR MANUFACTURING PROCESS AT THIS TIME.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

THE NEEDLE WAS INSERTED INTO THE INTERVERTEBRAL SPACE FOR ANESTHESIA SUB-ARACNOIDA THE MOMENT OF THE NEEDLE EXTRACTION THERE WAS DISCONNECTED CONNECTOR FROM THE NEEDLE, SO THE NEEDLE REMAINED INSERTED IN THE COLUMN THE NEEDLE WAS EXTRACTED WITH KLEMMER BECAUSE THE TAIL PROTRUDED A FEW MILLIMETERS FROM THE SKIN. INVOLVED PATIENT. INCIDENT THAT OCCURRED ON (B)(6) 2023. HAS THE DEVICE BEEN USED? YES. CONSEQUENCE: SPECIFIC INTERVENTION . SAMPLE AVAILABLE.

Description of Event or Problem · 0

THE NEEDLE WAS INSERTED INTO THE INTERVERTEBRAL SPACE FOR ANESTHESIA SUB-ARACNOIDA THE MOMENT OF THE NEEDLE EXTRACTION THERE WAS DISCONNECTED CONNECTOR FROM THE NEEDLE, SO THE NEEDLE REMAINED INSERTED IN THE COLUMN THE NEEDLE WAS EXTRACTED WITH KLEMMER BECAUSE THE TAIL PROTRUDED A FEW MILLIMETERS FROM THE SKIN. INVOLVED PATIENT: INCIDENT THAT OCCURRED ON (B)(6) 2023. HAS THE DEVICE BEEN USED? YES. CONSEQUENCE: SPECIFIC INTERVENTION. SAMPLE AVAILABLE.

Description of Event or Problem · 0

PLEASE VERIFY WHERE THE BREAK OCCURRED, DID A PIECE OF THE SPINAL NEEDLE BREAK OFF IN THE PATIENT? DID THE NEEDLE DISCONNECT FROM THE HUB? IS A PHOTO AVAILABLE? WAS ANY MEDICAL INTERVENTION REQUIRED? IT IS CONFIRMED THAT THE NEEDLE HAS DISCONNECTED FROM THE HUB AND HAS THEREFORE REMAINED WITHIN THE PATIENT'S SPINE. IT WAS NECESSARY FOR THE ANESTHESIOLOGIST WITH KLEMMER TO EXTRACT THE NEEDLE WHICH FORTUNATELY PROTRUDED A FEW MM. AS ALREADY SPECIFIED IN THE INCIDENT REPORT, THE SAMPLE IS AVAILABLE AND IS OBVIOUSLY CONTAMINATED WITH BLOOD AS SHOWN IN THE ATTACHED IMAGES. TOMORROW, WE WILL GIVE YOU FEEDBACK IN ORDER TO PREPARE FOR POSSIBLE WITHDRAWAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2296514 SPINAL SET 25GA 3-1/2IN NEEDLE, SPINAL, SHORT TERM MIA BECTON DICKINSON 2210027 00382904050849

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown