FDA Adverse Event Malfunction Summary report: N

CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ MEDIUM

MDR report key: 18393935 · Received December 26, 2023

Report

Report Number
2029046-2023-03070
Event Type
Malfunction
Date Received
December 26, 2023
Date of Event
November 27, 2023
Report Date
February 13, 2024
Manufacturer
BIOSENSE WEBSTER INC
Product Code
DYB
UDI-DI
10846835016277
PMA / PMN Number
K170997
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HAS BEEN REPORTED AS DISCARDED, THEREFORE NO PRODUCT INVESTIGATION CAN BE PERFORMED, AND THE CUSTOMER COMPLAINT CANNOT BE CONFIRMED. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE FINISHED DEVICE 60000197 NUMBER, AND NO INTERNAL ACTIONS RELATED TO THE COMPLAINT WERE FOUND DURING THE REVIEW. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY BIOSENSE WEBSTER, INC., OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, BIOSENSE WEBSTER, INC., OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

INITIALLY, IT WAS REPORTED THAT A HEMOSTATIC VALVE LEAK ISSUE OCCURRED. THE BIOSENSE WEBSTER, INC. PRODUCT ANALYSIS LAB RECEIVED THE DEVICE AND OBSERVED THAT THE HEMOSTATIC VALVE WAS BROKEN IN THE STAR SECTION. THE HEMOSTATIC VALVE LEAK ISSUE REMAINS ASSESSED AS MDR REPORTABLE. THE ADDITIONAL FINDING OF THE BROKEN HEMOSTATIC VALVE IN THE STAR SECTION WAS ALSO ASSESSED AS MDR REPORTABLE. THE AWARENESS DATE IS 22-JAN-2024. DEVICE EVALUATION DETAILS: THE DEVICE WAS RETURNED TO BIOSENSE WEBSTER (BWI) FOR EVALUATION. VISUAL INSPECTION AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE WERE PERFORMED FOLLOWING BWI PROCEDURES. VISUAL ANALYSIS REVEALED THAT THE HEMOSTATIC VALVE WAS BROKEN IN THE STAR SECTION. MICROSCOPIC EXAMINATION OF THE HEMOSTATIC VALVE SURFACE SHOWED STRESS MARKS ON THE OUTER DIAMETER. THE DAMAGE OBSERVED COULD BE RELATED TO THE INCORRECT INSERTION OF THE DILATOR INTO THE SHEATH. THE STRESS MARKS AND PHYSICAL DAMAGE OBSERVED SUGGEST THAT EXCESSIVE FORCE MANIPULATION WAS APPLIED; HOWEVER, THIS COULD NOT BE CONCLUSIVELY DETERMINED. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE FINISHED DEVICE 60000197 NUMBER, AND NO INTERNAL ACTIONS RELATED TO THE COMPLAINT WERE FOUND DURING THE REVIEW. THE ISSUE REPORTED BY THE CUSTOMER WAS CONFIRMED. THE ODP (OPTIMAL DEVICE PERFORMANCE GUIDE) CONTAINS THE FOLLOWING CAUTION: ALWAYS INSERT A DILATOR STRAIGHT INTO THE CENTER OF THE SHEATH¿S VALVE TO PREVENT DAMAGE TO THE VALVE. DO NOT INSERT A DILATOR AT AN ANGLE, AS DAMAGE TO THE SHEATH VALVE MAY OCCUR. AS PART OF THE QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THIS PRODUCT ISSUE WILL BE ADDRESSED THROUGH BWI'S QUALITY SYSTEM. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) ABLATION PROCEDURE WITH A CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ MEDIUM AND A THERE WAS RESISTANCE WHILE FLUSHING THE VIZIGO SHEATH. WHEN THE PHYSICIAN WENT TO FLUSH THE SHEATH, AIR AND BLOOD RETURNED (MINIMAL) WAS OBSERVED ON THE VIZIGO SHEATH'S HEMOSTATIC VALVE SECTION. WHEN THE VIZIGO SHEATH WAS REPLACED, THE ISSUE WAS RESOLVED. NO ADVERSE PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2266884 CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ MEDIUM INTRODUCER, CATHETER DYB BIOSENSE WEBSTER INC 60000197 10846835016277

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown