FDA Adverse Event Malfunction Summary report: N

TECNIS IOL

MDR report key: 18392330 · Received December 26, 2023

Report

Report Number
3012236936-2023-03276
Event Type
Malfunction
Date Received
December 26, 2023
Report Date
February 23, 2024
Manufacturer
AMO PUERTO RICO MFG. INC.
Product Code
HQL
PMA / PMN Number
P980040
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: SECTION D9: DEVICE AVAILABLE FOR EVALUATION: YES. SECTION D9: RETURNED TO MANUFACTURER ON: 01/03/2024. SECTION H3: DEVICE EVALUATED BY MANUFACTURER: YES. DEVICE EVALUATION: VISUAL INSPECTION REVEAL THAT THE LENS WAS CUT IN HALF AND COATED IN VISCOELASTIC RESIDUE. THE LENS WAS CLEANED AND DAMAGE ON THE SURFACE AND EDGE OF THE LENS WAS OBSERVED. NO FURTHER ISSUES WERE IDENTIFIED, AND NO FURTHER EVALUATION PERFORMED. CONCLUSION: THE COMPLAINT ISSUE "LENS DAMAGED" WAS IDENTIFIED DURING PRODUCT EVALUATION; HOWEVER, BASED ON THE COMPLAINT INVESTIGATION RESULTS THE COMPLAINT ISSUE COULD NOT BE CONFIRMED TO BE RELATED TO THE MANUFACTURING OR DESIGN PROCESS. THEREFORE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY OR PRODUCT MALFUNCTION. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

SECTION B3: DATE OF EVENT: UNKNOWN, NOT PROVIDED, BUT THE BEST ESTIMATE DATE IS BETWEEN 11/1/2023 AND 12/4/2023. SECTION H3: OTHER 81: THE DEVICE WAS NOT RETURNED FOR EVALUATION. THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORD, COMPLAINT TRENDING, AND RISK DOCUMENTATION FOR THIS DEVICE WILL BE PERFORMED. UPON COMPLETION OF THE REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION A SUPPLEMENTAL MEDWATCH WILL BE FILED. ATTEMPTS WERE MADE TO OBTAIN THE MISSING INFORMATION, HOWEVER TO DATE, NO RESPONSE HAD BEEN RECEIVED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE NON PRELOADED INTRAOCULAR LENS WAS CRACKED. NO OTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2269711 TECNIS IOL INTRAOCULAR LENS HQL AMO PUERTO RICO MFG. INC. ZCB00

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown