MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL
Report
- Report Number
- 2032227-2023-322008
- Event Type
- Malfunction
- Date Received
- December 26, 2023
- Date of Event
- November 5, 2021
- Report Date
- December 26, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000414344
- PMA / PMN Number
- P150001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. THE PUMP PASSED THE DISPLACEMENT TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF-TEST. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUMP. POWER MANAGEMENT GRAPH IS WITHIN SPECIFICATION. THE P-CAP LOCKS PROPERLY INTO THE RESERVOIR COMPARTMENT. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CRACKED KEYPAD OVERLAY, STAINED KEYPAD OVERLAY, PILLOWING KEYPAD OVERLAY, CRACKED CASE (BATTERY TUBE), SCRATCHED CASE AND SCRATCHED LCD WINDOW (MINOR SCRATCHES). FAILED BATT ALARM IN THE DETAILED TRACES ON (B)(6) 2021 17:26:37.000, (B)(6) 2021 17:29:17.000, (B)(6) 2021 17:32:22.000, (B)(6) 2021 17:36:57.000 AND (B)(6) 2021 17:39:03.000 WAS EXPECTED. NO BATTERY ANOMALIES NOTED. COSMETIC DAMAGE WAS CONFIRMED AT THE FRONT AND BACK OF PUMP. FAILED BATT TEST NOT CONFIRMED DUE TO EXPECTED BATTERY BEHAVIOR. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED THAT THE PUMP TURNS OFF AND ON AND DOES NOT RECOGNIZE THE BATTERIES WHEN YOU PUT A NEW ONE, THE BATTERY COVER.TROUBLESHOOTING WAS PERFORMED AND FOUND CUSTOMER STATES THE SELF TEST PASSED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE USING THE INSULIN PUMP AND IT WAS RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2264005 | MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1880 | HG59JWXZZ | 000000763000414344 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR | Unknown |