FDA Adverse Event Injury Summary report: N

RESERVOIR 3ML MMT-332A

MDR report key: 18391223 · Received December 26, 2023

Report

Report Number
2032227-2023-321968
Event Type
Injury
Date Received
December 26, 2023
Date of Event
November 21, 2023
Report Date
March 3, 2024
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
P150001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. THE INFORMATION WHICH HAS BEEN RECEIVED WAS NOT INCLUDED IN THE INITIAL REPORT. THE INFORMATION HAS BEEN PROVIDED IN SECTION B5, H6, H10 AND H11 WITH THIS REPORT.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

CUSTOMER REPORTED HAVING A LOW AND TREATING IT WITH GLUCAGON. PER CASE-2023-00984760, HE USED THE SENSORS FROM THE AFFECTED LOT F1323 FOR THE 780G G4S RECALL THAT COULD POTENTIALLY TRANSMIT INACCURATE SENSOR GLUCOSE READINGS. HE SAID HE WAS HAVING SENSOR UPDATING AND CHANGE SENSOR ALERTS. THE SENSORS WERE NOT LASTING 6-7 DAYS EVEN WITH A NEW TRANSMITTER. HE SAID DURING THE NIGHT ON (B)(6), 2023 (PER CURRENT CASE) HIS BLOOD GLUCOSE DROPPED AND HE LOST CONSCIOUSNESS AND HIS WIFE HAD TO CALL THE AMBULANCE AND THEY GAVE HIM THE GLUCAGON. PER HELPLINE, HE DID NOT MEASURE BLOOD GLUCOSE VALUES. USER ALLEGED OVER DELIVERY BECAUSE HE NEVER HAD THE SENSOR ISSUE BEFORE AND BECAUSE HIS LOT WAS A BAD LOT. PUMP WAS USED. PER CUSTOMER, SMARTGUARD WAS USED. HOWEVER, IF CUSTOMER WAS HAVING SENSOR UPDATING ISSUES, THEN SMARTGUARD WOULD NOT BE ACTIVE. TROUBLESHOOTING WAS DONE FOR THE LOW.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA WITH A BLOOD GLUCOSE VALUE OF 2.8MMOL/L AT THE TIME OF EVENT WITH THE SYMPTOMS OF UNCONSCIOUS , THE CUSTOMER WAS TREATED WITH GLUCAGON. TROUBLESHOOTING WAS PERFORMED AND IT WAS VERIFIED THAT PUMP WAS UTILIZED WITHIN 48 HOURS OF THE EVENT ALONG WITH ACTIVE AUTO MODE FEATURE AND WAS REPORTED AUTO MODE/SMARTGUARD WAS AUTOMATICALLY DELIVERING INSULIN AT THE TIME OF LOW BLOOD GLUCOSE EVENT AND CUSTOMER ALLEGES THE POSSIBLE PUMP OVER DELIVERY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. THE CUSTOMER WILL CONTINUE THE USE OF THE DEVICE AND THE RESERVOIR WILL NOT BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1490014 RESERVOIR 3ML MMT-332A PUMP, INFUSION FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-332A

Patients

Seq Age Sex Outcome Treatment
1 62 YR Unknown Required Intervention OZP-MMT-7040-SNSR, UNOMED SET