FDA Adverse Event Malfunction Summary report: N

ZENITH TX2 DISSECTION ENDOVASCULAR GRAFT STRAIGHT COMPONENT

MDR report key: 18388843 · Received December 22, 2023

Report

Report Number
3002808486-2023-00301
Event Type
Malfunction
Date Received
December 22, 2023
Date of Event
December 13, 2023
Report Date
January 24, 2024
Manufacturer
WILLIAM COOK EUROPE
Product Code
MIH
UDI-DI
10827002474489
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURERS REF# (B)(4). BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN OR UNAVAILABLE. G4) PMA/510(K): P180001. INVESTIGATION IS STILL IN PROGRESS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

MANUFACTURERS REF# (B)(4). SUMMARY OF INVESTIGATIONAL FINDINGS: A PATIENT OF UNDISCLOSED GENDER AND AGE UNDERWENT A REPAIR OF A CHRONIC DISSECTION WITH AN EXISTING NON-COOK PROXIMAL STENT GRAFT IN WHICH A ZDEG-P-32-142-PF-US (COMPLAINT DEVICE) WAS USED. THERE WAS NO NOTABLE TORTUOSITY OR ANGULATION. THE DEVICE DEPLOYED FINE AND AS INTENDED, HOWEVER THE SAFETY WIRE RELEASE WAS EXTREMELY DIFFICULT TO RELEASE. THE SAME DEVICE WAS PLACED DISTALLY WITH NO ISSUES ON RELEASE OR DEPLOYMENT. REVIEW OF THE DEVICE HISTORY RECORD FOUND THAT ALL DISCOVERED NON-CONFORMANCES WERE PROPERLY DISPOSITIONED BEFORE RELEASE AND NO INDICATION THAT THE DEVICE WAS PRODUCED OUTSIDE OF SPECIFICATION. THE COMPLAINT DEVICE WAS NOT RETURNED FOR THE INVESTIGATION AND NO IMAGING WAS PROVIDED. BASED ON THE RECEIVED INFORMATION IT HAS NOT BEEN POSSIBLE TO DETERMINE THE CAUSE OF DIFFICULT RELEASE OF THE GREEN TRIGGER-WIRE RELEASE MECHANISM. RELEVANT PERSONNEL HAVE BEEN INFORMED OF THIS COMPLAINT AND COOK WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS. IT IS NOTED THAT THE ZDEG WAS USED IN COMBINATION WITH A NON-COOK DEVICE, WHICH IS OUTSIDE OF INTENDED USE FOR THIS TYPE OF DEVICE. COOK MEDICAL WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: A PATIENT OF UNDISCLOSED GENDER AND AGE UNDERWENT A REPAIR OF A CHRONIC DISSECTION WITH EXISTING PROXIMAL STENT GRAFT IN WHICH THE ZENITH TX2 DISSECTION ENDOVASCULAR GRAFT STRAIGHT COMPONENT, G47448, WAS USED. THE GRAFT WAS BEING DEPLOYED INSIDE AN EXISTING SURGICAL GRAFT FROM ANOTHER MANUFACTURER. THERE WAS NO NOTABLE TORTUOSITY OR ANGULATION. THE DEVICE DEPLOYED FINE AND AS INTENDED, HOWEVER THE SAFETY WIRE RELEASE WAS EXTREMELY DIFFICULT TO RELEASE. ANOTHER UNIT OF THE SAME, G47448- LOT E4468992, WAS PLACED DISTALLY WITH NO ISSUES ON RELEASE OR DEPLOYMENT. PATIENT OUTCOME: THE COMPLAINANT DID NOT REPORT ANY ADVERSE EFFECTS TO THE PATIENT DUE TO THIS OCCURRENCE.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2264960 ZENITH TX2 DISSECTION ENDOVASCULAR GRAFT STRAIGHT COMPONENT MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH WILLIAM COOK EUROPE E4369124 10827002474489

Patients

Seq Age Sex Outcome Treatment
1 Unknown