HD AUTOCLAVABLE CAMERA HEAD
Report
- Report Number
- 3002808148-2023-20012
- Event Type
- Malfunction
- Date Received
- December 22, 2023
- Date of Event
- November 24, 2023
- Report Date
- February 9, 2024
- Manufacturer
- SHIRAKAWA OLYMPUS CO., LTD.
- Product Code
- FET
- UDI-DI
- 04953170228834
- PMA / PMN Number
- K955404
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
G1: MANUFACTURING CONTACT FACILITY NAME: SHIRAKAWA OLYMPUS CO., LTD. TO DATE, THE DEVICE HAS NOT BEEN RETURNED TO OLYMPUS FOR EVALUATION. THE INVESTIGATION IS ONGOING AND FOLLOW-UP WITH THE USER FACILITY IS CURRENTLY BEING PERFORMED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION OR IF ANY ADDITIONAL INFORMATION IS PROVIDED BY THE USER FACILITY.
THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION AND THE CUSTOMER'S ALLEGATION OF DISPLAYED IMAGE WAS CONFIRMED. DEVICE EVALUATION FOUND THAT THE REPORTED ISSUE WAS DUE TO A FAULTY CABLE. ADDITIONALLY, OTHER EVALUATION FINDINGS INCLUDE THE FOLLOWING: SCOPE DID NOT ROTATE SMOOTHLY. DEVICE HISTORY RECORD (DHR) REVIEW: A DHR WAS REVIEWED AND NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE WERE FOUND. INSTRUCTIONS FOR USE (IFU): ¿SHOULD ANY IRREGULARITY BE OBSERVED, DO NOT USE THE CAMERA HEAD; CONTACT OLYMPUS.¿ THE MOST PROBABLE CAUSE OF THE COMPLAINT IS CAUSE TRACED TO COMPONENT FAILURE. OLYMPUS WILL CONTINUE TO MONITOR THE FIELD PERFORMANCE OF THIS DEVICE.
THE CUSTOMER REPORTED TO OLYMPUS, THE HIGH DEFINITION AUTOCLAVABLE CAMERA HEAD IMAGE QUALITY WAS POOR, AND IMAGE DISPLACED. THERE WERE NO REPORTS OF PATIENT HARM.
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION RESULTS. IN ADDITION, A CORRECTION TO THIS REPORT WAS MADE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2288444 | HD AUTOCLAVABLE CAMERA HEAD | CAMERA HEAD | FET | SHIRAKAWA OLYMPUS CO., LTD. | OTV-S7PROH-HD-12E | 04953170228834 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |