FDA Adverse Event Malfunction Summary report: N

OT PING METER

MDR report key: 1838622 · Received September 19, 2010

Report

Report Number
2939301-2010-08175
Event Type
Malfunction
Date Received
September 19, 2010
Report Date
September 12, 2010
Manufacturer
LIFESCAN, INC.
Product Code
MDS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT; 510(K)# IS K082590.

Additional Manufacturer Narrative · 1

FOLLOW UP # 1/SUPPLEMENTAL REPORT TEXT 11/03/2010. THE LAY USER/PATIENT'S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER INVOLVED IN THIS CASE HAS PASSED TESTING WITH NO FAULTS FOUND. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A SECOND FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO THE SYSTEM HYDROPNEUMATIC THAT WAS LEAKING INTO THE HYDRO TRAY. NO INJURY WAS REPORTED.

Description of Event or Problem · 1

THE LAY USER/PATIENT CONTACTED LIFESCAN ALLEGING THE METER POWERS OFF DURING USE. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT PING METER GLUCOSE MONITORING SYS/KIT MDS LIFESCAN, INC.

Patients

Seq Age Sex Outcome Treatment
1