FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 18383963 · Received December 22, 2023

Report

Report Number
1221359-2023-01792
Event Type
Malfunction
Date Received
December 22, 2023
Date of Event
December 8, 2023
Report Date
February 14, 2024
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

FIRST TEST LOT INFORMATION: D4-LOT:225360 D4-EXPIRATION:25AUG2024 D4-UDI:(B)(4) SECOND TEST LOT INFORMATION: D4-LOT:223499 D4-EXPIRATION:03AUG2024 D4-UDI:(B)(4) FIRST LOT INVESTIGATION RESULTS: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 225360 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160/ LOT: 225360, TEST BASE PART NUMBER 195-430H/ LOT: 221803. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 225360 SHOWED THAT THE COMPLAINT RATE IS (B)(4). A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 225360 SHOWED THAT THE COMPLAINT RATE IS 0(B)(4)%. ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED; HOWEVER, IT COULD HAVE POSSIBLY BEEN RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE. SECOND LOT INVESTIGATION RESULTS: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 223499 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160/ LOT: 223499, TEST BASE PART NUMBER 195-430H/ LOT: 218296. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 223499 SHOWED THAT THE COMPLAINT RATE IS (B)(4). A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 223499 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD HAVE POSSIBLY BEEN RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED

Additional Manufacturer Narrative · 0

FIRST TEST LOT INFORMATION: D4-LOT:225360. D4-EXPIRATION:25AUG2024. D4-UDI:(B)(4). SECOND TEST LOT INFORMATION: D4-LOT:223499. D4-EXPIRATION:03AUG2024. D4-UDI:(B)(4). THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED CONFLICTING RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST KIT PERFORMED ON (B)(6)2023 ON AN UNKNOWN SAMPLE TYPE. ON (B)(6)2023, THE FIRST TEST TAKEN (LOT NUMBER: 225360) GENERATED A POSITIVE RESULT, AND THE SECOND TEST (LOT NUMBER: 223499), PERFORMED ON THE SAME DAY, GENERATED A NEGATIVE RESULT. THE CONSUMER INDICATED THEY WERE ASYMPTOMATIC. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED CONFLICTING RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST KIT PERFORMED ON (B)(6) 2023 ON AN UNKNOWN SAMPLE TYPE. ON (B)(6) 2023, THE FIRST TEST TAKEN (LOT NUMBER: 225360) GENERATED A POSITIVE RESULT, AND THE SECOND TEST (LOT NUMBER: 223499), PERFORMED ON THE SAME DAY, GENERATED A NEGATIVE RESULT. THE CONSUMER INDICATED THEY WERE ASYMPTOMATIC. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2296795 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. UNKNOWN 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 39 YR Male