FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 18381917 · Received December 22, 2023

Report

Report Number
1645337-2023-15041
Event Type
Injury
Date Received
December 22, 2023
Date of Event
November 13, 2023
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001218
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON (B)(6) 2024, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. MENTOR ALSO RECEIVED ADDITIONAL INFORMATION INDICATING THAT THE PATIENT'S IMPLANTS WERE REPLACED WITH THE FOLLOWING: (LEFT) 225CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE CATALOG: 3501630 LOT: 9905682 SN: (B)(6) AND (RIGHT) 225CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE CATALOG: 3501630 LOT: 9905682 SN: (B)(6). ON (B)(6) 2024, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE SAL SMOOTH RND DIAP 225CC BREAST IMPLANT HAD A CREASE/FOLD ON THE ANTERIOR VIEW EXTENDING TO THE POSTERIOR VIEW. IN ADDITION, A TEAR WAS NOTED ON THE POSTERIOR VIEW WITHIN THE CREASE MEASURING APPROXIMATELY 0.5 CM. A MICROSCOPIC EXAMINATION WAS PERFORMED AND INSTRUMENT DAMAGE WAS NOT FOUND ON THE AREA OF THE TEAR. THE EVALUATION DETERMINED THAT THE POSSIBLE CAUSE OF THE RUPTURE WAS CONSISTENT WITH NORMAL WEAR. DEFLATION CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT IT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. CREATING WRINKLES OR FOLDS SHOULD BE AVOIDED DURING IMPLANTATION OR OTHER PROCEDURES AS IT CAN RESULT IN DEFLATION AT A LATER TIME. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. A SECOND PRODUCT WAS RECEIVED (LOT-9672076). NO ADVERSE EVENTS WERE REPORTED FOR THIS CONCOMITANT (CONTRALATERAL) DEVICE, THEREFORE NO FURTHER ANALYSIS IS REQUIRED. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW.

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: MATERIAL RUPTURE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH TWO 225CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE BREAST PROSTHESES. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA PHYSICAL EXAMINATION, WITH LEFT BREAST IMPLANT DEFLATION. AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2023.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2051445 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 9672076 00081317001218

Patients

Seq Age Sex Outcome Treatment
1 67 YR Female Required Intervention