Report Number
3014704491-2023-00841
Event Type
Malfunction
Date Received
December 21, 2023
Date of Event
October 8, 2023
Report Date
February 13, 2024
Manufacturer
BD (SUZHOU)
Product Code
FOZ
UDI-DI
00382903830831
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

1. DHR/BHR REVIEW(LOT#3171581): 1)THIS BATCH OF PRODUCTS WERE ASSEMBLED AT INTIMA II AUTO LINE 2 IN JULY 2023, AND PACKAGED AT R240 PACKAGE LINE IN JULY 2023. WORK ORDER QUANTITY WAS (B)(4) EA. 2)REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MEET THE PRODUCT SPECIFICATIONS. 3)REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 2. NO DEFECTIVE SAMPLES AND PHOTOS HAVE BEEN RECEIVED FOR THE COMPLAINT. 3. 45PSI LEAKAGE TEST IS CARRIED OUT ON THE RETAINED SAMPLE OF THIS BATCH, AND NO LEAKAGE IS FOUND. PLEASE REFER TO ATTACHMENT FOR THE TEST REPORT. 4. NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. CONCLUSION(S): NO ABNORMALITY IS FOUND ON PROCESS AND RETAINED SAMPLE, AND NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. AS THE DAMAGE STATE OF THE COMPLAINED SAMPLE CANNOT BE CONFIRMED, THE ROOT CAUSE OF THE LEAKAGE CANNOT BE DETERMINED. THE PLANT WILL CONTINUE TO PAY ATTENTION TO SUCH DEFECTS. H3 OTHER TEXT : SEE NARRATIVE.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: (B)(4) 2023 PATIENT WAS ADMITTED TO THE HOSPITAL FOR A COMPRESSION FRACTURE OF THE THORACIC SPINE AND WAS GIVEN PAIN MANAGEMENT THAT REQUIRED THE USE OF A CLOSED IV INDWELLING NEEDLE TO ADMINISTER FLUIDS; THE STEEL NEEDLE CONNECTION WAS FOUND TO BE LEAKING AND WAS REPLACED WITHOUT HARM TO THE PATIENT.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION PROVIDED.