STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2023-32336
- Event Type
- Injury
- Date Received
- December 21, 2023
- Report Date
- February 12, 2024
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ADDITIONAL, CHANGED, AND/OR CORRECTED DATA. LABORATORY ANALYSIS SUMMARY: THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION WAS RECEIVED ON JANUARY 11, 2024, WITH LOT NUMBER: 3212839. BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: DEFLATION: NOT OBSERVED OPENINGS DURING THE ANALYSIS. ADDITIONAL OBSERVATIONS: NO OTHER OBSERVATION OBSERVED. NO FURTHER ACTIONS ARE REQUIRED SINCE NO MANUFACTURING ISSUE IS OBSERVED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. THE DEVICE HAS BEEN EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. THE DEVICE HAS BEEN EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 289088 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 3212839 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female | Required Intervention |