FDA Adverse Event Injury Summary report: N

DAVINCI XI

MDR report key: 18373130 · Received December 21, 2023

Report

Report Number
2955842-2023-21606
Event Type
Injury
Date Received
December 21, 2023
Date of Event
December 2, 2023
Report Date
December 2, 2023
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

BASED ON THE CURRENT INFORMATION PROVIDED, THE SURGEON CONVERTED THE PROCEDURE TO OPEN SURGERY DUE TO ISSUES WITH REPEATED UNRECOVERABLE FAULTS. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS ABLE TO REPRODUCE THE ISSUE AND REPLACED THE REDUNDANT POWER TRAY ASSEMBLY. THE SYSTEM WAS TESTED AND WAS READY FOR USE. ISI DID RECEIVE THE REDUNDANT POWER TRAY ASSEMBLY TO PERFORM FAILURE ANALYSIS (FA). FA WAS ABLE TO CONFIRM AND REPRODUCE THE CUSTOMER REPORTED COMPLAINT. THIS REDUNDANT POWER TRAY ASSEMBLY WAS TESTED ON AN IN-HOUSE SYSTEM. THE PROGRAM AND SYSTEM STARTED AT NORMAL MODE WITH NO ERROR TRIGGERING AT START UP. RAN 10 POWER CYCLES AND IT PASSED, HOWEVER, DURING IDLE TEST, ERROR 86 WAS TRIGGERED AFTER 15 MINUTES. CHECKING ON SYSTEM LOGS THE ADC VOLTAGE VALUE WAS SLIGHTLY BELOW THE MINIMUM RANGE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED DIAPHRAGMATIC HERNIA REPAIR SURGICAL PROCEDURE THAT AN UNRECOVERABLE FAULT OCCURRED. PRIOR TO CALLING, THE CALLER POWER CYCLED THE SYSTEM WITH NO CHANGE. ERROR 86 WAS OBSERVED IN THE LOGS. THE INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) WALKED THE CALLER THROUGH A HARD POWER CYCLE OF THE VISION SIDE CART (VSC) AND THE ERROR DID NOT PERSIST. THE CUSTOMER CONTACTED THE TSE AGAIN AFTER THE PROCEDURE AND STATED THAT THE ERROR OCCURRED AGAIN, AND ANOTHER POWER CYCLE WAS PERFORMED, BUT THE ERROR PERSISTED. THE SURGEON ELECTED TO CONVERT THE PROCEDURE TO OPEN SURGERY DUE TO THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
304543 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-45 N/A 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 Unknown DA VINCI INSTRUMENTS AND ACCESSORIES.