DAVINCI XI
Report
- Report Number
- 2955842-2023-21606
- Event Type
- Injury
- Date Received
- December 21, 2023
- Date of Event
- December 2, 2023
- Report Date
- December 2, 2023
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110898
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
BASED ON THE CURRENT INFORMATION PROVIDED, THE SURGEON CONVERTED THE PROCEDURE TO OPEN SURGERY DUE TO ISSUES WITH REPEATED UNRECOVERABLE FAULTS. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS ABLE TO REPRODUCE THE ISSUE AND REPLACED THE REDUNDANT POWER TRAY ASSEMBLY. THE SYSTEM WAS TESTED AND WAS READY FOR USE. ISI DID RECEIVE THE REDUNDANT POWER TRAY ASSEMBLY TO PERFORM FAILURE ANALYSIS (FA). FA WAS ABLE TO CONFIRM AND REPRODUCE THE CUSTOMER REPORTED COMPLAINT. THIS REDUNDANT POWER TRAY ASSEMBLY WAS TESTED ON AN IN-HOUSE SYSTEM. THE PROGRAM AND SYSTEM STARTED AT NORMAL MODE WITH NO ERROR TRIGGERING AT START UP. RAN 10 POWER CYCLES AND IT PASSED, HOWEVER, DURING IDLE TEST, ERROR 86 WAS TRIGGERED AFTER 15 MINUTES. CHECKING ON SYSTEM LOGS THE ADC VOLTAGE VALUE WAS SLIGHTLY BELOW THE MINIMUM RANGE.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED DIAPHRAGMATIC HERNIA REPAIR SURGICAL PROCEDURE THAT AN UNRECOVERABLE FAULT OCCURRED. PRIOR TO CALLING, THE CALLER POWER CYCLED THE SYSTEM WITH NO CHANGE. ERROR 86 WAS OBSERVED IN THE LOGS. THE INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) WALKED THE CALLER THROUGH A HARD POWER CYCLE OF THE VISION SIDE CART (VSC) AND THE ERROR DID NOT PERSIST. THE CUSTOMER CONTACTED THE TSE AGAIN AFTER THE PROCEDURE AND STATED THAT THE ERROR OCCURRED AGAIN, AND ANOTHER POWER CYCLE WAS PERFORMED, BUT THE ERROR PERSISTED. THE SURGEON ELECTED TO CONVERT THE PROCEDURE TO OPEN SURGERY DUE TO THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 304543 | DAVINCI XI | VISION SIDE SYSTEM | NAY | INTUITIVE SURGICAL, INC | 381121-45 | N/A | 00886874110898 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES. |