HAMILTON-C3
Report
- Report Number
- 3001421318-2023-12059
- Event Type
- Malfunction
- Date Received
- December 21, 2023
- Date of Event
- August 3, 2022
- Report Date
- August 29, 2024
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- PMA / PMN Number
- K201306
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE COMPLAINT HAS BEEN REOPENED AND REVIEWED ACCORDING TO FDA FORM 483 INSPECTIONAL OBSERVATION EMS #2, EOBS2 FROM THE FDA INSPECTION CONDUCTED BETWEEN JULY 17 TO JULY 21, 2022 AT THE EMS AND BONADUZ SITES. A DETAILED INVESTIGATION WAS PERFORMED BY AN EXPERT FROM THE TECHNICAL SERVICE: SINCE THE COMPLAINT IN QUESTION WAS SUBMITTED TO HAMILTON MEDICAL AG OVER 1 YEAR AGO, NO ATTEMPT WILL BE MADE TO OBTAIN ADDITIONAL INFORMATION. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT WHILE THE VENTILATOR WAS UNDERGOING PREVENTIVE MAINTENANCE. THE ROOT CAUSE OF THE PROBLEM WAS A DEFECTIVE/UNTIGHT NEBULIZER VALVE IN THE MIXER ASSEMBLY. THERE WAS NO PATIENT OR USER HARM. HAMILTON MEDICAL AG COMPLAINT NUMBER: (B)(4). FOLLOW-UP 1 - CORRECTED INFORMATION: UDI RELATED DATA QUALITY UPDATES ONLY. FIELD D4 WAS UPDATED WITH FULL UDI INFORMATION AS REQUESTED.
THE COMPLAINT HAS BEEN REOPENED AND REVIEWED ACCORDING TO FDA FORM 483 INSPECTIONAL OBSERVATION EMS #2, EOBS2 FROM THE FDA INSPECTION CONDUCTED BETWEEN JULY 17 TO JULY 21, 2022 AT THE EMS AND BONADUZ SITES. A DETAILED INVESTIGATION WAS PERFORMED BY AN EXPERT FROM THE TECHNICAL SERVICE: SINCE THE COMPLAINT IN QUESTION WAS SUBMITTED TO HAMILTON MEDICAL AG OVER 1 YEAR AGO, NO ATTEMPT WILL BE MADE TO OBTAIN ADDITIONAL INFORMATION. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT WHILE THE VENTILATOR WAS UNDERGOING PREVENTIVE MAINTENANCE. THE ROOT CAUSE OF THE PROBLEM WAS A DEFECTIVE/UNTIGHT NEBULIZER VALVE IN THE MIXER ASSEMBLY. THERE WAS NO PATIENT OR USER HARM.
NEBULIZER FAILS TO OPEN.
NEBULIZER FAILS TO OPEN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 277417 | HAMILTON-C3 | HAMILTON-C3 | CBK | HAMILTON MEDICAL AG | 160005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |