FDA Adverse Event Malfunction Summary report: N

HF SENSOR DELIVERY SYSTEM

MDR report key: 18367359 · Received December 20, 2023

Report

Report Number
3004936110-2023-01882
Event Type
Malfunction
Date Received
December 20, 2023
Date of Event
November 27, 2023
Report Date
February 13, 2024
Manufacturer
ST. JUDE MEDICAL, INC.
Product Code
MOM
UDI-DI
05414734509237
PMA / PMN Number
P100045
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING SENSOR WAS REVIEWED AND FOUND OUTSIDE OF THE TOLERANCE FOR CM MEAN. THE CAUSE IS INCONCLUSIVE AT THIS TIME AND IS UNDER INVESTIGATION. THIS REPORTED EVENT WAS INVESTIGATED FOR POSSIBLE INACCURATE READING. BASED ON THE INFORMATION GIVEN AND BACKEND LOG ANALYSIS, IT WAS CONFIRMED THAT THE DEVICE WAS OPERATING WITHIN THE EXPECTED FREQUENCY RANGE OF 30-37.5 MHZ. THE SENSOR WAS OPERATING AT 33.6, 33.8, AND 33.7 MHZ DURING THE REVIEW OF THE APPLICABLE READING(S). A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THAT NO NON-CONFORMANCES WERE IDENTIFIED THAT ARE RELATED TO THE REPORTED EVENT. PER THE IFU, RIGHT HEART CATHETERIZATION IS REQUIRED FOR SYSTEM BASELINE (MEAN PRESSURE) CALIBRATION AND MAY BE NEEDED TO RECALIBRATE THE BASELINE (MEAN PRESSURE) IN ORDER TO CONTINUE TO USE THE SYSTEM.

Description of Event or Problem · 0

A RIGHT HEART CATHETERIZATION WAS PERFORMED PRIMARILY FOR REASONS UNRELATED TO THE PRESSURE SENSOR. THE SENSOR PRESSURE READINGS WERE COMPARED TO THE PRESSURE READINGS FROM THE CATHETER. THE SENSOR WAS RECALIBRATED AND THE BASELINE WAS DECREASE BY 13MMHG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2118815 HF SENSOR DELIVERY SYSTEM SYSTEM, HEMODYNAMIC, IMPLANTABLE MOM ST. JUDE MEDICAL, INC. CM2000 8915510 05414734509237

Patients

Seq Age Sex Outcome Treatment
1 72 YR Male Required Intervention