FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 18361103 · Received December 19, 2023

Report

Report Number
2955842-2023-21456
Event Type
Malfunction
Date Received
December 19, 2023
Date of Event
November 27, 2023
Report Date
November 30, 2023
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874119792
PMA / PMN Number
K214095
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE MARYLAND BIPOLAR FORCEPS INSTRUMENT TO PERFORM FAILURE ANALYSIS. THE MARYLAND BIPOLAR FORCEPS INSTRUMENT WAS ANALYZED AND FOUND TO HAVE DAMAGED CONDUCTOR WIRE INSULATION AT THE DISTAL END. THE WIRES WERE EXPOSED. THE INSTRUMENT WAS PLACED AND DRIVEN ON AN IN-HOUSE SYSTEM. THE INSTRUMENT PASSED THE RECOGNITION AND ENGAGEMENT TESTS AND MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS. THE GRIPS OPENED AND CLOSED PROPERLY. THE INSTRUMENT PASSED ELECTRICAL CONTINUITY AND ENERGY DELIVERY TESTS. THERE WAS NO THERMAL DAMAGE AND NO MISSING MATERIAL.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) HAS RECEIVED THE INSTRUMENT; HOWEVER, FAILURE ANALYSIS HAS NOT COMPLETED THEIR INVESTIGATION. A FOLLOW-UP MDR WILL BE SUBMITTED WHEN FAILURE ANALYSIS HAS COMPLETED THEIR INVESTIGATION.

Description of Event or Problem · 0

REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE MARYLAND BIPOLAR FORCEPS HAD A DEEP CUT IN THE BLACK ENERGY CORD. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2206072 ENDOWRIST MARYLAND BIPOLAR FORCEPS NAY INTUITIVE SURGICAL, INC 471172-17 K11230629 0467 00886874119792

Patients

Seq Age Sex Outcome Treatment
1 Unknown DA VINCI INSTRUMENTS AND ACCESSORIES