FDA Adverse Event Malfunction Summary report: N

BARD® URINE METER WITH BAG

MDR report key: 18357591 · Received December 19, 2023

Report

Report Number
1018233-2023-09058
Event Type
Malfunction
Date Received
December 19, 2023
Date of Event
November 30, 2023
Report Date
January 9, 2024
Manufacturer
C.R. BARD, INC. (COVINGTON) -1018233
Product Code
KNX
UDI-DI
00801741033117
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED ISSUE IS INCONCLUSIVE AS NO SAMPLE WAS RETURNED FOR EVALUATION. A POTENTIAL ROOT CAUSE FOR THIS FAILURE COULD BE DUE TO "INCORRECT ASSEMBLY OPERATION OF ESTANE GLUE APPLICATION IN THE URINE METER ADAPTER AND URINE METER / DRAINAGE BAG ASSEMBLY.". THE DEVICE HISTORY RECORD WAS REVIEWED AND FOUND NOTHING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING: "BARD® EZ-LOK® SAMPLING PORT (INDICATED BY THE BLUE STEM IN THE PORT) ACCEPTS LUER LOCK (FIG. 1A) OR SLIP TIP SYRINGES (FIG. 1B). 1. OCCLUDE DRAINAGE TUBING A MINIMUM OF 3 INCHES BELOW THE SAMPLING PORT BY KINKING THE TUBING UNTIL URINE IS VISIBLE UNDER THE ACCESS SITE. 2. SWAB SURFACE OF BARD® EZ-LOK® SAMPLING PORT WITH ANTISEPTIC WIPE. (FIG. 2) 3. USING ASEPTIC TECHNIQUE, POSITION THE SYRINGE IN THE CENTER OF THE SAMPLING PORT. PRESS THE SYRINGE FIRMLY AND TWIST GENTLY TO ACCESS THE SAMPLING PORT. (FIG. 3) 4. SLOWLY ASPIRATE URINE SAMPLE INTO SYRINGE AND REMOVE SYRINGE FROM SAMPLE PORT. 5. UNKINK TUBING IF NECESSARY AND TRANSFER URINE SPECIMEN INTO SPECIMEN CUP OR FOLLOW HOSPITAL PROTOCOL. DISCARD SYRINGE ACCORDING TO HOSPITAL PROTOCOL. 6. FOLLOW ESTABLISHED HOSPITAL PROTOCOL FOR SPECIMEN LABELING AND TRANSPORT TO LAB. STERILE UNLESS PACKAGE IS OPENED OR DAMAGED, EXCEPT FOR ANY INDIVIDUALLY PACKAGED COMPONENTS WITHIN THE TRAY WHICH ARE NOT LABELED AS STERILE. THESE COMPONENTS ARE NOT TERMINALLY STERILIZED. CAUTION: FEDERAL (U.S.A.) LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN. SINGLE USE ONLY. DO NOT RESTERILIZE. FOR UROLOGICAL USE ONLY. WARNING: ON CATHETER, DO NOT USE OINTMENTS OR LUBRICANTS HAVING A PETROLATUM BASE. THEY WILL DAMAGE THE CATHETER AND MAY CAUSE BALLOON TO BURST. WARNING: AFTER USE, THIS PRODUCT MAY BE A POTENTIAL BIOHAZARD. HANDLE AND DISPOSE OF IN ACCORDANCE WITH ACCEPTED MEDICAL PRACTICES AND APPLICABLE LOCAL, STATE AND FEDERAL LAWS AND REGULATIONS. VISUALLY INSPECT THE PRODUCT FOR ANY IMPERFECTIONS OR SURFACE DETERIORATION PRIOR TO USE. IF PACKAGE IS OPENED OR IF ANY IMPERFECTION OR SURFACE DETERIORATION IS OBSERVED, DO NOT USE. PLEASE CONSULT PRODUCT LABEL AND INSERT FOR ANY INDICATIONS, CONTRAINDICATIONS, HAZARDS, WARNINGS, CAUTIONS AND DIRECTIONS FOR USE. PROPER TECHNIQUES FOR URINARY CATHETER INSERTION ¿ PERFORM HAND HYGIENE IMMEDIATELY BEFORE AND AFTER INSERTION. ¿ INSERT URINARY CATHETERS USING ASEPTIC TECHNIQUE AND STERILE EQUIPMENT. ¿ USE THE SMALLEST FOLEY CATHETER POSSIBLE, CONSISTENT WITH GOOD DRAINAGE. ¿ DOCUMENT THE INDICATIONS FOR CATHETER INSERTION, DATE AND TIME OF CATHETER INSERTION, INDIVIDUAL WHO INSERTED CATHETER, AND DATE AND TIME OF CATHETER REMOVAL IN PATIENT RECORD. PROPER TECHNIQUES FOR URINARY CATHETER MAINTENANCE. ¿ SECURE THE FOLEY CATHETER, USE THE STATLOCK® FOLEY STABILIZATION DEVICE IF PROVIDED. ¿ MAINTAIN A CLOSED DRAINAGE SYSTEM BY UTILIZING PRE-CONNECTED, SEALED CATHETER-TUBING JUNCTIONS. ¿ MAINTAIN UNOBSTRUCTED URINE FLOW AND KEEP THE CATHETER AND COLLECTION TUBE FREE FROM KINKING. ¿ KEEP THE COLLECTION BAG BELOW THE LEVEL OF THE BLADDER OR HIPS AT ALL TIMES. ¿ EMPTY THE COLLECTION BAG REGULARLY (E.G., PRIOR TO TRANSPORT) USING A SEPARATE, CLEAN COLLECTION CONTAINER FOR EACH PATIENT. ¿ ROUTINE HYGIENE (E.G., CLEANSING OF THE MEATAL SURFACE DURING DAILY BATHING OR SHOWERING) IS APPROPRIATE. ¿ LEAVE FOLEY CATHETER IN PLACE ONLY AS LONG AS NEEDED". H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. H3 OTHER TEXT : THE DEVICE WAS NOT RETURNED.

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. H3 OTHER TEXT : THE DEVICE WAS NOT RETURNED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE FOLEY CATHETER WOULD NOT DRAIN. NO DIFFICULTY WITH INSERTION BUT WHEN PRIMARY REGISTER NURSE ATTEMPTED TO DRAIN FOLEY CATHETER FROM DEPENDENT DRAIN BAG (NOT CLEAR MEASURED COLLECTION UNIT) THE SEAL ON THE WHITE SIDE AND CLEAR SIDE OF THE BAG WAS FUSED AND WOULD NOT ALLOW FOR DRAINAGE FROM DRAIN BAG. FOLEY REMAINED IN PLACE UNTIL DELIVERY UPON WHICH TIME IT WAS REMOVED WHEN PATIENT BEGAN PUSHING. NO HARM TO PATIENT OR BABY AS FOLEY WAS IN PLACE FOR A SHORT PERIOD OF TIME (LOT#NGHV3874). PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 28NOV2023, CUSTOMER REPORTED THAT FOLEY CATHETER HAD A DEFECT (LOT#NGHU2373). CERVICAL DILATION BALLOON PLACED ON (B)(6) 2023 AT 1833. CERVICAL DILATION BALLOON FOUND TO BE RUPTURED AND DEFLATED UPON ASSESSMENT. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 30NOV2023, CUSTOMER REPORTED TWO COMPLAINTS ABOUT FOLEY CATHETERS. BOTH COMPLAINTS WERE REGARDING A FOLEY BAG BEING FUSED TOGETHER (CLEAR PLASTIC PART AND WHITE PLASTIC PART) WHICH WOULD NOT ALLOW URINE TO FILL THE ENTIRE BAG (PCN#-153204 - LOT#NGHV3874 AND PCN#153204 - LOT#NGHV3874).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE FOLEY CATHETER WOULD NOT DRAIN. NO DIFFICULTY WITH INSERTION BUT WHEN PRIMARY REGISTER NURSE ATTEMPTED TO DRAIN FOLEY CATHETER FROM DEPENDENT DRAIN BAG (NOT CLEAR MEASURED COLLECTION UNIT) THE SEAL ON THE WHITE SIDE AND CLEAR SIDE OF THE BAG WAS FUSED AND WOULD NOT ALLOW FOR DRAINAGE FROM DRAIN BAG. FOLEY REMAINED IN PLACE UNTIL DELIVERY UPON WHICH TIME IT WAS REMOVED WHEN PATIENT BEGAN PUSHING. NO HARM TO PATIENT OR BABY AS FOLEY WAS IN PLACE FOR A SHORT PERIOD OF TIME (LOT#: NGHV3874). PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 28NOV2023, CUSTOMER REPORTED THAT FOLEY CATHETER HAD A DEFECT (LOT#: NGHU2373). CERVICAL DILATION BALLOON PLACED ON (B)(6) 2023 AT 1833. CERVICAL DILATION BALLOON FOUND TO BE RUPTURED AND DEFLATED UPON ASSESSMENT. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 30NOV2023, CUSTOMER REPORTED TWO COMPLAINTS ABOUT FOLEY CATHETERS. BOTH COMPLAINTS WERE REGARDING A FOLEY BAG BEING FUSED TOGETHER (CLEAR PLASTIC PART AND WHITE PLASTIC PART) WHICH WOULD NOT ALLOW URINE TO FILL THE ENTIRE BAG (PCN#:153204 - LOT#: NGHV3874 AND PCN#: 153204 - LOT#: NGHV3874).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1497356 BARD® URINE METER WITH BAG URINE METER WITH BAG KNX C.R. BARD, INC. (COVINGTON) -1018233 NGHV3874 00801741033117

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other