SENHANCE
Report
- Report Number
- 3007593944-2023-00018
- Event Type
- Malfunction
- Date Received
- December 19, 2023
- Date of Event
- August 30, 2022
- Report Date
- October 30, 2024
- Manufacturer
- ASENSUS SURGICAL, INC.
- Product Code
- NAY
- UDI-DI
- 00815440020781
- PMA / PMN Number
- K171120
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
BASED ON THE COMPLAINT DETAILS, NO ADVERSE EVENT OCCURRED TO A PATIENT OR USER. THE EVENT OCCURRED ON 30-AUG-2022, AND ASENSUS SURGICAL WAS MADE AWARE 01-SEP-2022. THIS EVENT OCCURRED IN THE UNITED STATES AT (B)(6). THERE WAS A <15 MINUTE DELAY TO THE SURGICAL CASE WHICH IS ACCEPTABLE IN THE DEVICE RISK DOCUMENTATION. THERE WAS NO IMPACT TO THE PATIENT, AND THE OPERATION WAS COMPLETED WITH SENHANCE. THE COMPLAINANT STATED THAT WHILE USING ULTRASONIC WITH 5 MM BIPOLAR INSTRUMENTATION, THE SURGEON GRABBED UTERINE VESSELS TO CAUTERIZE WITH BIPOLAR ENERGY WHICH APPEARED TO DELIVER TO THE INTENDED TISSUE BUT THEN A HIGH PITCH NOTICE SOUNDED FROM THE CONMED 5000 GENERATOR. SOON AFTER GRABBING THE UTERINE VESSELS, A LOW TONE SOUNDED FROM THE GENERATOR. THE GENERATOR WAS THEN EXCHANGED FROM THE CONMED 5000 TO THE COVIDIEN FT10. STILL AFTER PRESSING THE PEDAL AND GRABBING THE VESSEL THERE WAS NO CONVERSATION FROM THE BIPOLAR INSTRUMENT. AFTER THE CASE ON (B)(6) 2022 AN INSPECTION WAS DONE ON THE BIPOLAR INSTRUMENT WITH A SODIUM CHLORIDE SOAKED 4 X 4 CLOTH CONNECTED TO THE CONMED 5000 AND THERE WAS INTERMITTENT ENERGY DELIVERY WHILE CLOSING THE JAWS ALONG WITH A SPARK WHILE DELIVERING ENERGY. DURING THE SECOND TEST, THE POWER CABLE WAS SWITCHED OUT BUT STILL SIMILAR EFFECT SHOWING MINIMAL INTERMITTENT ENERGY DELIVERY. THEN IN THIRD TEST, THE INSTRUMENT INSERT WAS SWITCHED OUT WITH ANOTHER BIPOLAR INSTRUMENT INSERT AND THIS ACTION APPEARED TO DELIVER ENERGY BETTER. UPON FURTHER INSPECTION, THERE APPEARED TO BE DAMAGE TO THE BLUE INSULATION OF THE TIP OF THE BIPOLAR INSTRUMENT. ASENSUS SURGICAL REPORTED A SIMILAR COMPLAINT: (B)(4) ON 30-NOV-2023, WHICH LED THE COMPANY TO REEVALUATE PREVIOUS COMPLAINTS WITH SIMILAR OCCURRENCES OF INSTRUMENT SPARKING. IN COMPLAINT: (B)(4), THE ROOT CAUSE WAS ATTRIBUTED TO WEAR AND TEAR OF THE INSULATION PART OF THE INSTRUMENT. MOREOVER, THREE SIMILAR COMPLAINTS: (B)(4) WERE REPORTED PRIOR TO THIS COMPLAINT THAT ALSO REPORTED SPARKING. THEIR ROOT CAUSE WAS ATTRIBUTED TO EITHER HUMAN ERROR DURING BIPOLAR ENERGY APPLICATION OR FACTORS ASSOCIATED WITH INSTRUMENT WEAR AND TEAR. ACCORDING TO THE INVESTIGATION, THE REPORTED ISSUE OCCURRED WITH A 5 MM BIPOLAR LARGE GRASPING FORCEPS, P/N: X0005146, LOT: 987880, UPON WHICH A DHR REVIEW WAS PERFORMED WITH NO ANOMALOUS FINDINGS. AS THE INSTRUMENT WAS MANUFACTURED IN APRIL 2021, MULTIPLE USAGES CAN BE EXPECTED IN THIS TIMEFRAME. THE ROOT CAUSE OF THIS ISSUE CAN, THEREFORE, BE ATTRIBUTED TO WEAR AND TEAR. THIS USE FATIGUE CAN RESULT IN DAMAGE TO THE BIPOLAR INSTRUMENT INSULATION. NO ANOMALIES WERE FOUND IN THE DHR, AND MANUFACTURING DEFECTS CAN BE EXCLUDED. FURTHERMORE, IFU FOR BIPOLAR SURGICAL INSTRUMENTS, UM-001-00046 ADVISES THE FOLLOWING: INSPECT INSTRUMENTS AND CABLES FOR DAMAGE BEFORE EACH USE, ESPECIALLY THE INSULATION OF LAPAROSCOPIC/ENDOSCOPIC INSTRUMENTS. THIS MAY BE DONE VISUALLY UNDER MAGNIFICATION OR WITH A HIGH VOLTAGE INSULATION TESTING DEVICE. INSULATION FAILURES MAY RESULT IN BURNS OR OTHER INJURIES TO THE PATIENT OR OPERATOR. THE USE LIFE OF THE SENHANCE BIPOLAR INSTRUMENTS IS TYPICALLY DETERMINED BY FACTORS SUCH AS WEAR AND DAMAGE DUE TO USE AND THE NATURE OF HANDLING BETWEEN USES. DO NOT USE DEVICES THAT ARE DAMAGED, EXCESSIVELY WORN, OR NOT WORKING PROPERLY IN ORDER TO PREVENT PATIENT OR OPERATOR INJURY. ALTHOUGH THERE WAS NO IMPACT ON THE PATIENT AND INSTRUMENT WEAR AND TEAR IS CONSIDERED, THE MALFUNCTION OF THE DEVICE IS COMPATIBLE WITH THE DEFINITION FOR A SERIOUS INCIDENT AND MEETS THE REPORTING CRITERIA SINCE IT MIGHT HAVE LED TO DAMAGE TO THE SURROUNDING TISSUES CAUSED BY THE HEAT. ADDITIONALLY, IF THE MALFUNCTION WERE TO RECUR, THERE COULD BE SERIOUS INJURY TO THE PATIENT. THEREFORE, THE DECISION IS TO REPORT THIS EVENT. FOLLOWUP #1: CORRECTED CATALOG AND MODEL NUMBER FIELDS THAT WERE INADVERTENTLY ENTERED INTO THE SERIAL NUMBER FIELD. PER FDA REQUEST, INCLUDED THE FULL (UDI) FOR THIS DEVICE ASSOCIATED WITH THE EVENT, SINCE ASENSUS PREVIOUSLY PROVIDED ONLY THE GTIN AND LOT NUMBERS.
BASED ON THE COMPLAINT DETAILS, NO ADVERSE EVENT OCCURRED TO A PATIENT OR USER. THE EVENT OCCURRED ON 30-AUG-2022, AND ASENSUS SURGICAL WAS MADE AWARE (B)(6) 2022. THIS EVENT OCCURRED IN THE (B)(6) IN WINTER PARK, FL. THERE WAS A <15 MINUTE DELAY TO THE SURGICAL CASE WHICH IS ACCEPTABLE IN THE DEVICE RISK DOCUMENTATION. THERE WAS NO IMPACT TO THE PATIENT, AND THE OPERATION WAS COMPLETED WITH SENHANCE. THE COMPLAINANT STATED THAT WHILE USING ULTRASONIC WITH 5 MM BIPOLAR INSTRUMENTATION, THE SURGEON GRABBED UTERINE VESSELS TO CAUTERIZE WITH BIPOLAR ENERGY WHICH APPEARED TO DELIVER TO THE INTENDED TISSUE BUT THEN A HIGH PITCH NOTICE SOUNDED FROM THE CONMED 5000 GENERATOR. SOON AFTER GRABBING THE UTERINE VESSELS, A LOW TONE SOUNDED FROM THE GENERATOR. THE GENERATOR WAS THEN EXCHANGED FROM THE CONMED 5000 TO THE COVIDIEN FT10. STILL AFTER PRESSING THE PEDAL AND GRABBING THE VESSEL THERE WAS NO CONVERSATION FROM THE BIPOLAR INSTRUMENT. AFTER THE CASE ON (B)(6) 2022 AN INSPECTION WAS DONE ON THE BIPOLAR INSTRUMENT WITH A SODIUM CHLORIDE SOAKED 4 X 4 CLOTH CONNECTED TO THE CONMED 5000 AND THERE WAS INTERMITTENT ENERGY DELIVERY WHILE CLOSING THE JAWS ALONG WITH A SPARK WHILE DELIVERING ENERGY. DURING THE SECOND TEST, THE POWER CABLE WAS SWITCHED OUT BUT STILL SIMILAR EFFECT SHOWING MINIMAL INTERMITTENT ENERGY DELIVERY. THEN IN THIRD TEST, THE INSTRUMENT INSERT WAS SWITCHED OUT WITH ANOTHER BIPOLAR INSTRUMENT INSERT AND THIS ACTION APPEARED TO DELIVER ENERGY BETTER. UPON FURTHER INSPECTION, THERE APPEARED TO BE DAMAGE TO THE BLUE INSULATION OF THE TIP OF THE BIPOLAR INSTRUMENT. ASENSUS SURGICAL REPORTED A SIMILAR COMPLAINT (B)(4) ON (B)(6) 2023, WHICH LED THE COMPANY TO REEVALUATE PREVIOUS COMPLAINTS WITH SIMILAR OCCURRENCES OF INSTRUMENT SPARKING. IN COMPLAINT (B)(4) THE ROOT CAUSE WAS ATTRIBUTED TO WEAR AND TEAR OF THE INSULATION PART OF THE INSTRUMENT. MOREOVER, THREE SIMILAR COMPLAINTS (B)(4) WERE REPORTED PRIOR TO THIS COMPLAINT THAT ALSO REPORTED SPARKING. THEIR ROOT CAUSE WAS ATTRIBUTED TO EITHER HUMAN ERROR DURING BIPOLAR ENERGY APPLICATION OR FACTORS ASSOCIATED WITH INSTRUMENT WEAR AND TEAR. ACCORDING TO THE INVESTIGATION, THE REPORTED ISSUE OCCURRED WITH A 5 MM BIPOLAR LARGE GRASPING FORCEPS, P/N X0005146, LOT 987880, UPON WHICH A DHR REVIEW WAS PERFORMED WITH NO ANOMALOUS FINDINGS. AS THE INSTRUMENT WAS MANUFACTURED IN APRIL 2021, MULTIPLE USAGES CAN BE EXPECTED IN THIS TIMEFRAME. THE ROOT CAUSE OF THIS ISSUE CAN, THEREFORE, BE ATTRIBUTED TO WEAR AND TEAR. THIS USE FATIGUE CAN RESULT IN DAMAGE TO THE BIPOLAR INSTRUMENT INSULATION. NO ANOMALIES WERE FOUND IN THE DHR, AND MANUFACTURING DEFECTS CAN BE EXCLUDED. FURTHERMORE, IFU FOR BIPOLAR SURGICAL INSTRUMENTS, UM-001-00046 ADVISES THE FOLLOWING: INSPECT INSTRUMENTS AND CABLES FOR DAMAGE BEFORE EACH USE, ESPECIALLY THE INSULATION OF LAPAROSCOPIC/ENDOSCOPIC INSTRUMENTS. THIS MAY BE DONE VISUALLY UNDER MAGNIFICATION OR WITH A HIGH VOLTAGE INSULATION TESTING DEVICE. INSULATION FAILURES MAY RESULT IN BURNS OR OTHER INJURIES TO THE PATIENT OR OPERATOR. THE USE LIFE OF THE SENHANCE BIPOLAR INSTRUMENTS IS TYPICALLY DETERMINED BY FACTORS SUCH AS WEAR AND DAMAGE DUE TO USE AND THE NATURE OF HANDLING BETWEEN USES. DO NOT USE DEVICES THAT ARE DAMAGED, EXCESSIVELY WORN, OR NOT WORKING PROPERLY IN ORDER TO PREVENT PATIENT OR OPERATOR INJURY. ALTHOUGH THERE WAS NO IMPACT ON THE PATIENT AND INSTRUMENT WEAR AND TEAR IS CONSIDERED, THE MALFUNCTION OF THE DEVICE IS COMPATIBLE WITH THE DEFINITION FOR A SERIOUS INCIDENT AND MEETS THE REPORTING CRITERIA SINCE IT MIGHT HAVE LED TO DAMAGE TO THE SURROUNDING TISSUES CAUSED BY THE HEAT. ADDITIONALLY, IF THE MALFUNCTION WERE TO RECUR, THERE COULD BE SERIOUS INJURY TO THE PATIENT. THEREFORE, THE DECISION IS TO REPORT THIS EVENT.
ON 30-AUG-2022, IN THE UNITED STATES AT (B)(6), IT WAS REPORTED IN (B)(4) THAT WHILE USING ULTRASONIC WITH 5 MM BIPOLAR INSTRUMENTATION, THE SURGEON GRABBED UTERINE VESSELS TO CAUTERIZE WITH BIPOLAR ENERGY WHICH APPEARED TO DELIVER TO THE INTENDED TISSUE BUT THEN A HIGH PITCH NOTICE SOUNDED FROM THE CONMED 5000 GENERATOR. SOON AFTER GRABBING THE UTERINE VESSELS, A LOW TONE SOUNDED FROM THE GENERATOR. THE GENERATOR WAS THEN EXCHANGED FROM THE CONMED 5000 TO THE COVIDIEN FT10. STILL AFTER PRESSING THE PEDAL AND GRABBING THE VESSEL THERE WAS NO CONVERSATION FROM THE BIPOLAR INSTRUMENT. AFTER THE CASE ON (B)(6) 2022, AN INSPECTION WAS DONE ON THE BIPOLAR INSTRUMENT WITH A SODIUM CHLORIDE SOAKED 4 X 4 CLOTH CONNECTED TO THE CONMED 5000 AND THERE WAS INTERMITTENT ENERGY DELIVERY WHILE CLOSING THE JAWS ALONG WITH A SPARK WHILE DELIVERING ENERGY. DURING THE SECOND TEST, THE POWER CABLE WAS SWITCHED OUT BUT STILL SIMILAR EFFECT SHOWING MINIMAL INTERMITTENT ENERGY DELIVERY. THEN IN THIRD TEST, THE INSTRUMENT INSERT WAS SWITCHED OUT WITH ANOTHER BIPOLAR INSTRUMENT INSERT AND THIS ACTION APPEARED TO DELIVER ENERGY BETTER. UPON FURTHER INSPECTION, THERE APPEARED TO BE DAMAGE TO THE BLUE INSULATION OF THE TIP OF THE BIPOLAR INSTRUMENT. THERE WAS NO IMPACT TO THE PATIENT OR THE SURGICAL CASE, AND THE OPERATION WAS COMPLETED WITH THE SENHANCE SYSTEM. ASENSUS SURGICAL REPORTED A SIMILAR COMPLAINT: (B)(4) ON 30-NOV-2023, WHICH LED THE COMPANY TO REEVALUATE PREVIOUS COMPLAINTS WITH SIMILAR OCCURRENCES OF INSTRUMENT SPARKING. IN COMPLAINT: (B)(4), THE ROOT CAUSE WAS ATTRIBUTED TO WEAR AND TEAR OF THE INSULATION PART OF THE INSTRUMENT. MOREOVER, THREE SIMILAR COMPLAINTS (B)(4) WERE REPORTED PRIOR TO THIS COMPLAINT THAT ALSO REPORTED SPARKING. THEIR ROOT CAUSE WAS ATTRIBUTED TO EITHER HUMAN ERROR DURING BIPOLAR ENERGY APPLICATION OR FACTORS ASSOCIATED WITH INSTRUMENT WEAR AND TEAR. ALTHOUGH THERE WAS NO IMPACT ON THE PATIENT IN THIS PARTICULAR EVENT, THE MALFUNCTION OF THE DEVICE IS COMPATIBLE WITH THE DEFINITION FOR A SERIOUS INCIDENT AND MEETS THE REPORTING CRITERIA SINCE IT MIGHT HAVE LED TO DAMAGE TO THE SURROUNDING TISSUES CAUSED BY THE HEAT. ADDITIONALLY, IF THE MALFUNCTION WERE TO RECUR, THERE COULD BE SERIOUS INJURY TO THE PATIENT. THEREFORE, THE DECISION IS TO REPORT THIS EVENT.
ON (B)(6) 2022, IN THE (B)(6) HOSPITAL IN WINTER PARK, FL, IT WAS REPORTED IN (B)(4) THAT WHILE USING ULTRASONIC WITH 5 MM BIPOLAR INSTRUMENTATION, THE SURGEON GRABBED UTERINE VESSELS TO CAUTERIZE WITH BIPOLAR ENERGY WHICH APPEARED TO DELIVER TO THE INTENDED TISSUE BUT THEN A HIGH PITCH NOTICE SOUNDED FROM THE CONMED 5000 GENERATOR. SOON AFTER GRABBING THE UTERINE VESSELS, A LOW TONE SOUNDED FROM THE GENERATOR. THE GENERATOR WAS THEN EXCHANGED FROM THE CONMED 5000 TO THE COVIDIEN FT10. STILL AFTER PRESSING THE PEDAL AND GRABBING THE VESSEL THERE WAS NO CONVERSATION FROM THE BIPOLAR INSTRUMENT. AFTER THE CASE ON (B)(6) 2022 AN INSPECTION WAS DONE ON THE BIPOLAR INSTRUMENT WITH A SODIUM CHLORIDE SOAKED 4 X 4 CLOTH CONNECTED TO THE CONMED 5000 AND THERE WAS INTERMITTENT ENERGY DELIVERY WHILE CLOSING THE JAWS ALONG WITH A SPARK WHILE DELIVERING ENERGY. DURING THE SECOND TEST, THE POWER CABLE WAS SWITCHED OUT BUT STILL SIMILAR EFFECT SHOWING MINIMAL INTERMITTENT ENERGY DELIVERY. THEN IN THIRD TEST, THE INSTRUMENT INSERT WAS SWITCHED OUT WITH ANOTHER BIPOLAR INSTRUMENT INSERT AND THIS ACTION APPEARED TO DELIVER ENERGY BETTER. UPON FURTHER INSPECTION, THERE APPEARED TO BE DAMAGE TO THE BLUE INSULATION OF THE TIP OF THE BIPOLAR INSTRUMENT.THERE WAS NO IMPACT TO THE PATIENT OR THE SURGICAL CASE, AND THE OPERATION WAS COMPLETED WITH THE SENHANCE SYSTEM.ASENSUS SURGICAL REPORTED A SIMILAR COMPLAINT (B)(4) ON (B)(6) 2023, WHICH LED THE COMPANY TO REEVALUATE PREVIOUS COMPLAINTS WITH SIMILAR OCCURRENCES OF INSTRUMENT SPARKING. IN COMPLAINT (B)(4) THE ROOT CAUSE WAS ATTRIBUTED TO WEAR AND TEAR OF THE INSULATION PART OF THE INSTRUMENT. MOREOVER, THREE SIMILAR COMPLAINTS (B)(4) WERE REPORTED PRIOR TO THIS COMPLAINT THAT ALSO REPORTED SPARKING. THEIR ROOT CAUSE WAS ATTRIBUTED TO EITHER HUMAN ERROR DURING BIPOLAR ENERGY APPLICATION OR FACTORS ASSOCIATED WITH INSTRUMENT WEAR AND TEAR.ALTHOUGH THERE WAS NO IMPACT ON THE PATIENT IN THIS PARTICULAR EVENT, THE MALFUNCTION OF THE DEVICE IS COMPATIBLE WITH THE DEFINITION FOR A SERIOUS INCIDENT AND MEETS THE REPORTING CRITERIA SINCE IT MIGHT HAVE LED TO DAMAGE TO THE SURROUNDING TISSUES CAUSED BY THE HEAT. ADDITIONALLY, IF THE MALFUNCTION WERE TO RECUR, THERE COULD BE SERIOUS INJURY TO THE PATIENT. THEREFORE, THE DECISION IS TO REPORT THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1350558 | SENHANCE | SENHANCE SURGICAL SYSTEM | NAY | ASENSUS SURGICAL, INC. | X0005146 | 987880 | 00815440020781 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |