HF SENSOR DELIVERY SYSTEM
Report
- Report Number
- 3004936110-2023-01812
- Event Type
- Injury
- Date Received
- December 18, 2023
- Date of Event
- December 7, 2023
- Report Date
- March 7, 2024
- Manufacturer
- ST. JUDE MEDICAL, INC.
- Product Code
- MOM
- UDI-DI
- 05414734509237
- PMA / PMN Number
- P100045
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
NO DEVICE ANALYSIS RESULTS ARE AVAILABLE BECAUSE THE DEVICE REMAINS IMPLANTED. A REVIEW OF THE INFORMATION PROVIDED WAS PERFORMED, AND THE SENSOR WAS WITHIN THE TOLERANCE FOR CM MEAN; THEREFORE, THE REPORTED MALFUNCTION IS UNCONFIRMED. THIS REPORTED EVENT WAS INVESTIGATED FOR POSSIBLE INACCURATE READING. BASED ON THE INFORMATION GIVEN AND BACKEND LOG ANALYSIS, IT WAS CONFIRMED THAT THE DEVICE WAS OPERATING WITHIN THE EXPECTED FREQUENCY RANGE OF 30-37.5 MHZ. THE SENSOR WAS OPERATING AT 35.67-MHHZ, 35.71-MHZ, AND 35.51-MHZ DURING THE REVIEW OF THE APPLICABLE READING(S). A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THAT NO NON-CONFORMANCES WERE IDENTIFIED THAT ARE RELATED TO THE REPORTED EVENT. PER THE IFU, RIGHT HEART CATHETERIZATION IS REQUIRED FOR SYSTEM BASELINE (MEAN PRESSURE) CALIBRATION AND MAY BE NEEDED TO RECALIBRATE THE BASELINE (MEAN PRESSURE) IN ORDER TO CONTINUE TO USE THE SYSTEM.
A RIGHT HEART CATHETERIZATION WAS PERFORMED TO CONFIRM THE VALIDITY OF THE PRESSURE SENSOR READINGS. THE SENSOR WAS RECALIBRATED, AND THE MEAN WAS DECREASED BY 5.4 MMHG. READINGS WERE OBSERVED UNDER THE MANUFACTURER'S PATIENT CARE NETWORK DATABASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2303114 | HF SENSOR DELIVERY SYSTEM | SYSTEM, HEMODYNAMIC, IMPLANTABLE | MOM | ST. JUDE MEDICAL, INC. | CM2000 | 8866792 | 05414734509237 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Female | Required Intervention |