FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® 9NC 0.109M PLUS BLOOD COLLECTION TUBES

MDR report key: 18351207 · Received December 18, 2023

Report

Report Number
9617032-2023-01813
Event Type
Malfunction
Date Received
December 18, 2023
Date of Event
November 22, 2023
Report Date
December 20, 2023
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
50382903630955
PMA / PMN Number
K013971
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: MATERIAL #: 363095. LOT/BATCH #: 3180412. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 20 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING AND NO ISSUES WERE OBSERVED RELATING TO OVERFILL AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE OVERFILL. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE H.10.

Additional Manufacturer Narrative · 0

(B)(6). H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH THE BD VACUTAINER® 9NC 0.109M PLUS BLOOD COLLECTION TUBES, TUBES WERE OVERFILLING. THERE WAS NO REPORT OF PATIENT IMPACT.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH THE BD VACUTAINER® 9NC 0.109M PLUS BLOOD COLLECTION TUBES, TUBES WERE OVERFILLING. THERE WAS NO REPORT OF PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
78138 BD VACUTAINER® 9NC 0.109M PLUS BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 3180412 50382903630955

Patients

Seq Age Sex Outcome Treatment
1 Unknown