MAMMOTOME CONTROL MODULE WITH SMARTVAC - 120V
Report
- Report Number
- 1527736-2007-08406
- Event Type
- Malfunction
- Date Received
- December 11, 2007
- Date of Event
- January 1, 2007
- Report Date
- October 29, 2007
- Manufacturer
- ETHICON ENDO SURGERY, INC (CINCINNATI)
- Product Code
- KNW
- PMA / PMN Number
- K991980
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVAL SUMMARY - BASED ON ANALYSIS RESULTS, THIS COMPLAINT IS NOW DETERMINED TO BE A MDR MALFUNCTION. THE ANALYSIS SITE CONFIRMED THE CUSTOMER COMPLAINT AND PERFORMED SERVICE TESTS AND FOUND THE WIRES PULLED OUT OF CONNECTOR ON THE IF BOARD, CONFIRMING THE CUSTOMERS COMPLAINT OF NOT RECOGNIZING THE HOLSTER BEING CONNECTED, AND TO CORRECT THIS COMPLAINT THE ANALYSIS SITE REPLACED THE HOLSTER: IF BOARD TO BEZEL. THE VSO WAS REPLACED DUE TO THE SITE FOUND THAT IT FAILED IN TESTING, THE MAIN HOUSING MOUNTS WERE BROKEN ON THE INSIDE AND THE SITE REPLACED THE MAIN HOUSING, TUBING BULKHEAD GASKET, AND THE BEZEL ROPE GASKET TO CORRECT THE MAIN HOUSING ISSUE,AND THE PUMP MOUNT SCREWS (4), THE PUMP MOUNTS (4), FENDER WASHERS (4), AND FLAT WASHER (4) WERE REPLACED DUE TO THE PUMP MOUNTS WERE DAMAGED. AFTER SERVICING THE UNIT PASSED ALL QA FUNCTIONAL TESTING. THE DEVICE HISTORY RECORD HAS BEEN REVIEWED AND HAS PASSED QA INSPECTION. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED.
IT WAS REPORTED THAT THE X-RAY TABLE WAS ACCIDENTALLY LOWERED ONTO THE CONTROL MODULE. THE CUSTOMER STATED THAT THERE IS NO VISIBLE DAMAGE TO CONTROL MODULE BUT IT WILL NOT RECOGNIZE HOLSTER BEING CONNECTED. NO PT CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAMMOTOME CONTROL MODULE WITH SMARTVAC - 120V | KNW | ETHICON ENDO SURGERY, INC (CINCINNATI) | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | PROBE| HOLSTER |