FDA Adverse Event Malfunction Summary report: N

MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MMOL/L

MDR report key: 18345199 · Received December 18, 2023

Report

Report Number
2032227-2023-318001
Event Type
Malfunction
Date Received
December 18, 2023
Date of Event
November 20, 2023
Report Date
June 14, 2024
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000734435
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE PUMP PASSED ALL FUNCTIONAL TESTING INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST, AND THE DAT TEST AT .08720 INCHES. SUCCESSFULLY DOWNLOADED THE TRACE AND HISTORY FILES USING THUMP AND CARELINK UPLOAD WAS SUCCESSFUL. NO UNDER-DELIVERY ANOMALY WAS NOTED DURING TESTING. THE PUMP WAS CUT OPEN TO PERFORM A VISUAL INSPECTION. NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE WAS FOUND ON THE ELECTRONIC ASSEMBLY (PCBA 1 AND PCBA 2) AND MOTOR ASSEMBLY. A SC1 CAP LOCKS INTO PLACE INSIDE THE RESERVOIR COMPARTMENT PROPERLY. THE FOLLOWING WERE NOTED DURING A PHYSICAL: SCRATCHED CASE. THE PUMP PASSED ALL FUNCTIONAL TESTING. UNDER-DELIVERY AND HIGH BG ANOMALIES ARE NOT CONFIRMED. THE PUMP HISTORY FILE LISTS FOUR (4) BOLUSES ON THE EVENT DATE, 11/20/2023, LISTED BELOW: (B)(6) 2023 13:25:38.000 NORMALBOLUSDELIVERED (220) BOLUSPROGRAMMINGMETHOD: BOLUSWIZARD (1). NORMALBOLUSAMOUNTPROGRAMMED: 120000 (12 U) BOLUSAMOUNTDELIVERED: 120000 (12 U) (B)(6) 2023 15:35:22.000 NORMALBOLUSDELIVERED (220) BOLUSPROGRAMMINGMETHOD: BOLUSWIZARD (1). NORMALBOLUSAMOUNTPROGRAMMED: 106500 (10.65 U) BOLUSAMOUNTDELIVERED: 106500 (10.65 U) (B)(6) 2023 16:07:55.000 NORMALBOLUSDELIVERED (220) BOLUSPROGRAMMINGMETHOD: BOLUSWIZARD (1). NORMALBOLUSAMOUNTPROGRAMMED: 126000 (12.6 U) BOLUSAMOUNTDELIVERED: 126000 (12.6 U) (B)(6) 2023 19:01:36.000 NORMALBOLUSDELIVERED (220) BOLUSPROGRAMMINGMETHOD: BOLUSWIZARD (1). NORMALBOLUSAMOUNTPROGRAMMED: 153250 (15.325 U) BOLUSAMOUNTDELIVERED: 153250 (15.325 U) PLEASE SEE BELOW FOR THE DAILY TOTAL OF ALL INSULIN DELIVERED ON THE EVENT DATE, (B)(6) 2023: (B)(6) 2023 00:00:00.000 DAILYTOTALSG670 (63). DAILYTOTALCOLLECTIONSTARTTIME: (B)(6) 2023 00:00:00.000. DAILYTOTALOFALLINSULINDELIVERED: 899250 (89.925 U). DAILYTOTALOFBASALINSULINDELIVERED: 393500 (39.35 U). DAILYTOTALOFBOLUSINSULINDELIVERED: 505750 (50.575 U). THERE WERE NO AUTO SUSPEND EVENTS ON THE EVENT DATE, 11/20/2023. THERE WERE NO USER SUSPEND EVENTS ON THE EVENT DATE, 11/20/2023. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED OUTSIDE THE UNITED STATES. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA WITH A BLOOD GLUCOSE VALUE OF 28 MMOL/L. THE CUSTOMER ALSO REPORTED THE PUMP WAS NOT DELIVERING INSULIN. THE CUSTOMER WAS TREATED WITH AN INSULIN PEN. THE CUSTOMER WAS DISCONNECTED FROM THE INSULIN PUMP AND CONTINUOUS GLUCOSE MONITORING. NO FURTHER PATIENT COMPLICATIONS WHERE REPORTED. THE CUSTOMER WILL DISCONTINUE USING THE INSULIN PUMP. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
919801 MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MMOL/L AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1885 NG3512967H 000000763000734435

Patients

Seq Age Sex Outcome Treatment
1 14 YR Unknown