FDA Adverse Event Malfunction Summary report: N

LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (STANDARD)

MDR report key: 1834477 · Received September 10, 2010

Report

Report Number
2024601-2010-00697
Event Type
Malfunction
Date Received
September 10, 2010
Date of Event
July 26, 2010
Report Date
July 26, 2010
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TAPER II. (B)(4). VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ALLERGAN HAS RECEIVED THE PRODUCT; HOWEVER, THE ANALYSIS HAS NOT BEEN COMPLETED AT THIS TIME. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."

Description of Event or Problem · 1

RECEIVED A VOICEMAIL MESSAGE FROM HEALTH PROFESSIONAL REQUESTING A RETURN KIT FOR AN EXPLANTED DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED. FURTHER FOLLOW-UP WITH THE HEALTH PROFESSIONAL NOTED A "RAPID DEVICE LEAK."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (STANDARD) LTI ALLERGAN NA 1762129

Patients

Seq Age Sex Outcome Treatment
1 23 YR