FDA Adverse Event Death Summary report: N

630G INSULIN PUMP MMT-1715KL 630G

MDR report key: 18344696 · Received December 18, 2023

Report

Report Number
2032227-2023-317974
Event Type
Death
Date Received
December 18, 2023
Date of Event
December 6, 2022
Report Date
January 11, 2024
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
UDI-DI
000000763000316655
PMA / PMN Number
P150001
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. INFORMATION PROVIDED IN THE INITIAL REPORT IN SECTION G3 WAS SUBMITTED WITH INCORRECT 'AWARE DATE'. THE AWARE DATE SHOULD BE CONSIDERED AS '11-DEC-2023'.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER HAD PASSED AWAY AT HOSPITAL ON (B)(6) 2022. THE CUSTOMER WAS TREATED IN AN HOSPITAL PRIOR TO THE REPORTED INCIDENT. THE CAUSE OF DEATH WAS DUE TO HEART STROKE AND HIGH BLOOD GLUCOSE. THE CUSTOMER¿S BLOOD GLUCOSE AT THE TIME OF THE EVENT WAS UNKNOWN. IT WAS UNKNOWN WHETHER THE CUSTOMER WAS WEARING THE INSULIN PUMP AT THE TIME OF DEATH. THE CUSTOMER WILL DISCONTINUE USING THE INSULIN PUMP AND IT WILL NOT BE RETURNED FOR PRODUCT ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
919768 630G INSULIN PUMP MMT-1715KL 630G AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1715KL HG5P2G0 000000763000316655

Patients

Seq Age Sex Outcome Treatment
1 79 YR Female Death FRN-MMT-332A-RSVR,UNOMED INF SET