LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM
Report
- Report Number
- 2024601-2010-00691
- Event Type
- Malfunction
- Date Received
- September 10, 2010
- Date of Event
- March 16, 2010
- Report Date
- July 26, 2010
- Manufacturer
- ALLERGAN
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS NOTED THE BAND TUBING WAS BRITTLE AND CONTAINED A SHARP OPENING WITH LEAKAGE. THE BAND TUBING WAS BROKEN WITH STRIATIONS CONSISTENT WITH A SURGICAL END CUT TO REMOVE THE DEVICE. THE DAMAGED PORT SEPTUM HAD PULLED MATERIAL WITH EVIDENCE OF CORING LIKELY TO HAVE BEEN CAUSED BY THE USE OF A CORING NEEDLE. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."
RECEIVED A VOICEMAIL MESSAGE FROM HEALTH PROFESSIONAL REQUESTING A RETURN KIT FOR AN EXPLANTED DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED. FURTHER FOLLOW-UP WITH THE HEALTH PROFESSIONAL NOTED, THE PORT WAS EXPLANTED DUE TO AN ALLEGED LEAK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM | LTI | ALLERGAN | NA | 1104585 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |