FDA Adverse Event Malfunction Summary report: N

INFINITI VISION SYSTEM OZIL

MDR report key: 1834406 · Received September 10, 2010

Report

Report Number
2028159-2010-01714
Event Type
Malfunction
Date Received
September 10, 2010
Date of Event
January 1, 2010
Report Date
August 11, 2010
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K082845
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A COMPANY SERVICE REPRESENTATIVE EXAMINED THE SYSTEM AND WAS UNABLE TO DUPLICATE THE REPORTED PROBLEM. A SYSTEM MESSAGE WAS CONFIRMED IN THE EVENT LOG. THE SOLENOID SPACERS WERE REPLACED AND SENT FOR EVALUATION. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THE SOLENOID SPACERS HAVE BEEN RECEIVED AND IN-HOUSE TESTING IS IN PROGRESS. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. (B)(4).

Description of Event or Problem · 1

ADVERSE EVENT(S): "NO PATIENT INJURY REPORTED" (NO CONSEQUENCES OR IMPACT TO PATIENT). PRODUCT PROBLEM(S): "SYSTEM SHUT DOWN" (LOSS OF POWER). A BIOMEDICAL ENGINEER REPORTED THE UNIT SHUT DOWN DURING THE CASE. THE SYSTEM WAS SWITCHED OUT AND THE CASE WAS COMPLETED WITHOUT INCIDENT. THERE WAS NO PATIENT INJURY REPORTED.