OCTRODE PERCUTANEOUS LEAD
Report
- Report Number
- 1627487-2010-01796
- Event Type
- Injury
- Date Received
- September 13, 2010
- Date of Event
- December 1, 2009
- Report Date
- December 4, 2009
- Manufacturer
- ADVANCED NEUROMODULATION SYSTEMS
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVALUATION: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. A REVIEW OF THE DHR FOUND A NONCONFORMANCE RELATED TO THE PRODUCT, HOWEVER, THE NONCONFORMANCE WAS IDENTIFIED AS A COSMETIC ISSUE AND DID NOT AFFECT PRODUCT INTEGRITY OR PRODUCT FUNCTIONALITY AND WAS APPROVED FOR USE. THE DHR ANOMALY IS NOT RELATED TO THE ALLEGED DEVICE FAILURE. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. THIS MDR IS BEING SUBMITTED PAST THE 30 DAYS REPORTING REQUIREMENT AS PART OF A RETROSPECTIVE REVIEW INITIATED IN RESPONSE TO AN FDA INSPECTION. A RETROSPECTIVE REVIEW OF THE COMPLAINT RECORD DETERMINED THAT ANS MISINTERPRETED THE MDR REGULATIONS IN THIS INSTANCE. ANS HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
THE PATIENT RECEIVED HIS SCS SYSTEM, WHICH INCLUDED A PERCUTANEOUS LEAD AND AN IPG, ON (B)(6)2009. IT WAS REPORTED, THE PATIENT COULD ONLY ADJUST THE STIMULATION ON THE LEFT SIDE. IN ADDITION, STIMULATION HAD ALSO BECOME INTERMITTENT. THE LEAD WAS SURGICALLY REPOSITIONED. AS NO PRODUCTS WERE EXPLANTED DURING THE REPOSITIONING, NONE WERE RETURNED TO THE MANUFACTURER FOR ANALYSIS. FOLLOW UP ON THE PATIENT FOUND NO FURTHER ISSUES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OCTRODE PERCUTANEOUS LEAD | SPINAL CORD STIMULATION LEAD | LGW | ADVANCED NEUROMODULATION SYSTEMS | 3183 | 2790290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |