EON IMPLANTABLE PULSE GENERATOR
Report
- Report Number
- 1627487-2010-01698
- Event Type
- Injury
- Date Received
- September 13, 2010
- Date of Event
- July 31, 2008
- Report Date
- July 31, 2008
- Manufacturer
- ADVANCED NEUROMODULATION SYSTEMS
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVAL: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECS AND NO ANOMALIES WERE FOUND. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. THIS MDR IS BEING SUBMITTED PAST THE 30 DAY REPORTING REQUIREMENT AS PART OF A RETROSPECTIVE REVIEW INITIATED IN RESPONSE TO AN FDA INSPECTION. A RETROSPECTIVE REVIEW OF THE COMPLAINT RECORD DETERMINED THAT ANS MISINTERPRETED THE MDR REGULATIONS IN THIS INSTANCE. ANS HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
THE PT RECEIVED HER SCS SYSTEM ON (B)(6)2007. IT WAS REPORTED THAT THE SCS SYSTEM HEATS UP WHEN STIMULATION IS ON. IN ADDITION, THE PT ALSO FEELS DISCOMFORT NEAR THE HEADER OF THE IPG. BOTH THE HEATING AND THE DISCOMFORT CEASE ONCE THE STIMULATION HAD BEEN DISABLED. THE IPG WAS REPLACED ON (B)(6)2008. FOLLOW UP ON THE PT FOUND THAT NO FURTHER ISSUES HAVE BEEN REPORTED. THE EXPLANTED IPG WAS RETURNED TO ANS FOR EVAL. NO ADDITIONAL INFO IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EON IMPLANTABLE PULSE GENERATOR | TOTALLY IMPLANTABLE PULSE GENERATOR | LGW | ADVANCED NEUROMODULATION SYSTEMS | 3716 | 96420 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |