FDA Adverse Event Injury Summary report: N

INFINITI VISION SYSTEM OZIL

MDR report key: 1833639 · Received September 8, 2010

Report

Report Number
2028159-2010-01664
Event Type
Injury
Date Received
September 8, 2010
Date of Event
August 6, 2010
Report Date
August 9, 2010
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K082845
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY SERVICE REPRESENTATIVE EXAMINED THE SYSTEM AND FOUND THE IV POLE FUNCTION WAS FINE. ALL THE SURGICAL MODES WERE TESTED AND WERE IN CORRECT OPERATION. A SYSTEM MESSAGE WAS FOUND IN THE EVENT LOG WHICH INDICATED AN OBSTRUCTION IN THE HANDPIECE OR TUBING. THE CUSTOMER IS DIRECTED TO CLEAN AND/OR REPLACE THE HANDPIECE OR TUBING BY THE SYSTEM MESSAGE. THE COMPANY SERVICE REPRESENTATIVE FOUND THE STAFF WAS MANUALLY LOWERING THE IV POLE TO REMOVE THE IRRIGATION BOTTLE, INSERT A NEW CASSETTE, SPIKE A NEW IRRIGATION BOTTLE AND PRIME THE SYSTEM. AFTER PRIMING IS COMPLETED THE IV POLE LOWERS TO THE PREVIOUS LEVEL USED AND NOT THE MEMORY SETTING. THE IV POLE ISSUE WAS DISCUSSED WITH THE CUSTOMER. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. (B)(4)

Description of Event or Problem · 1

ADVERSE EVENT(S): "POSTERIOR CAPSULE TEAR" (CAPSULE BAG TEAR, POSTERIOR). PRODUCT PROBLEM(S): "IV POLE CAME DOWN FROM 95 TO 35 ITSELF" (NO CODE AVAILABLE); "LOW ASPIRATION" (ASPIRATION ISSUE). THE NURSE REPORTED LOSS OF CONTROL OF THE I/V POLE AND ASPIRATION VACUUM. A POSTERIOR CAPSULE TEAR OCCURRED. ADD'L INFO RECEIVED FROM THE NURSE STATED THAT DURING SURGERY THERE WAS LOW TO NO ASPIRATION VACUUM. THE HANDPIECE WAS CHANGED AND THE SYSTEM WAS PRIMED AGAIN. THE IV POLE CAME DOWN FROM 95 TO 35 BY ITSELF AND THE PT SUFFERED A POSTERIOR CAPSULE TEAR AND LOSS OF VITREOUS. THE SURGERY WAS COMPLETED AND THE IOL WAS IMPLANTED. ADD'L INFO RECEIVED STATED, A CHAMBER COLLAPSE CONTRIBUTED TO THE POSTERIOR CAPSULE TEAR. THIS COMPLICATION WAS DUE TO A STRONG SURGE DURING SURGERY. THE IOL WAS IMPLANTED IN THE SULCUS. THE SURGERY WAS COMPLETED IN AN HOUR AND 22 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITI VISION SYSTEM OZIL PHACOFRAGMENTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CENTER INFINITI OZIL NA

Patients

Seq Age Sex Outcome Treatment
1 84 YR Other MIOCHOL