FDA Adverse Event Malfunction Summary report: N

HARMONIC ACE 36 CM

MDR report key: 1833435 · Received September 14, 2010

Report

Report Number
3005075853-2010-05258
Event Type
Malfunction
Date Received
September 14, 2010
Date of Event
August 19, 2010
Report Date
August 20, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). ADDED ADDITIONAL INFORMATION THE DEVICE WAS RECEIVED WITH THE BLADE DISCOLORED (BLUE) DUE TO HEAT GENERATED FROM CONTACT BETWEEN THE BLADE AND TISSUE PAD. THE DEVICE WILL STOP ACTIVATING, AND EITHER EMIT A SOLID TONE OR DISPLAY AN INSTRUMENT ERROR CODE 5 ON THE GENERATOR DISPLAY WHEN THE BLADE BECOMES DAMAGED. THE BLADE WAS FOUND TO BE BROKEN AT THE DISCOLORED (BLUE). PROBABLE CAUSE OF BLADE DISCOLORATION IS APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TITANIUM TIP OF THE SCALPEL WAS BROKEN. THE BROKEN PART DID NOT FALL INTO THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Description of Event or Problem · 1

CALLER STATES PATIENT TESTED 7.0 INR ON THE COAGUCHEK XS SYSTEM WHILE A COMPARISON LAB RETURNED AS 4.5 INR. PATIENT WAS TREATED WITH AN ORAL DOSE OF VITAMIN K BASED ON METER RESULT. PATIENT'S COUMADIN DOSE WAS ALSO HELD FOR ONE DAY. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT SYSTEM AND REPLACEMENT WAS SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HARMONIC ACE 36 CM INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. UNK G4T67P

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE