FLEXION PACING LEAD
Report
- Report Number
- 1035166-2010-00051
- Event Type
- Malfunction
- Date Received
- September 9, 2010
- Date of Event
- July 19, 2010
- Report Date
- September 9, 2010
- Manufacturer
- OSCOR INC.
- Product Code
- DTB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THE LEAD WAS SURGICALLY ABANDONED (CAPPED), THEREFORE IT WILL NOT BE RETURNED FOR ANALYSIS. AS A RESULT, THE REPORTED ALLEGATION CANNOT BE CONFIRMED. THE EVENT WILL BE RE-EVALUATED IF ADD'L INFO BECOMES AVAILABLE. THRESHOLD ELEVATION IS A RECOGNIZED CLINICAL EVENT REFERENCE IN THE DEVICE'S LABELING AND/OR REPORTED IN THE MEDICAL LITERATURE. THE DEVICE HISTORY RECORDS AND THE COMPLAINT FILES OF THE LEAD MODEL WERE REVIEWED. NO TREND OF THE SAME OR SIMILAR TYPE WAS FOUND OR INDICATED.
THE CUSTOMER REPORTED THIS ATRIAL LEAD WAS SURGICALLY ABANDONED (CAPPED) DUE TO THRESHOLD MEASUREMENTS GREATER THAN 3.5 V/1.0 MS. IMPEDANCE MEASUREMENTS WERE ALSO LOW AROUND 290 OHMS AND R-WAVES WERE MEASURED AT 3.6 MV. NO ADVERSE PT EFFECTS WERE REPORTED. THE LEAD WAS ACTIVELY IMPLANTED FOR APPROX 6 YRS, 5 MONTHS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLEXION PACING LEAD | RETRACTABLE SCREW-IN PACING LEAD | DTB | OSCOR INC. | 4017 | 2T3 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR |