FDA Adverse Event
Malfunction
Summary report: N
UT69 TABLE MOUNTED RADIATION SHIELDS
MDR report key: 1833208
·
Received September 9, 2010
Report
- Report Number
- MW5017368
- Event Type
- Malfunction
- Date Received
- September 9, 2010
- Date of Event
- July 14, 2010
- Report Date
- September 9, 2010
- Manufacturer
- MAVIG GMBH
- Product Code
- KPY
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE UT69TABLE SHIELD HAS A U PROFILE THAT COVERS THE SHIELDS SUPPORT BAR AND THE PROFILE EDGES WERE SHARP CAUSING PAPER CUT TYPE INJURIES ON THE INSTALLER. (B)(6) DID A QUALITY CHECK AND FOUND OUT THAT THE MANUFACTURE DID NOT BURR (SAND AND POLISH) THE EDGES WHICH CAUSED THIS ISSUE. (B)(6) HAD THE MANUFACTURE FIX THE ISSUE AND WE WENT THROUGH AND REPLACED ALL THE DEFECTIVE PROFILES WITH ONES THAT WERE BURRED AND SANDED SO THEY HAD A SMOOTH EDGE AND DID NOT CAUSE ANY MORE CUTS. THE MANUFACTURER OF THE U PROFILES DETECTED THAT THE PROBLEM CAME FROM A NON SUFFICIENTLY TRAINED EMPLOYEE, HAS SINCE INCREASED HIS QUALITY AND TRAINING PROCESS AND ADAPTED A LASER PROCESSED PRODUCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UT69 TABLE MOUNTED RADIATION SHIELDS | UT69 | KPY | MAVIG GMBH | UT6902-US |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |