FDA Adverse Event Malfunction Summary report: N

UT69 TABLE MOUNTED RADIATION SHIELDS

MDR report key: 1833208 · Received September 9, 2010

Report

Report Number
MW5017368
Event Type
Malfunction
Date Received
September 9, 2010
Date of Event
July 14, 2010
Report Date
September 9, 2010
Manufacturer
MAVIG GMBH
Product Code
KPY
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE UT69TABLE SHIELD HAS A U PROFILE THAT COVERS THE SHIELDS SUPPORT BAR AND THE PROFILE EDGES WERE SHARP CAUSING PAPER CUT TYPE INJURIES ON THE INSTALLER. (B)(6) DID A QUALITY CHECK AND FOUND OUT THAT THE MANUFACTURE DID NOT BURR (SAND AND POLISH) THE EDGES WHICH CAUSED THIS ISSUE. (B)(6) HAD THE MANUFACTURE FIX THE ISSUE AND WE WENT THROUGH AND REPLACED ALL THE DEFECTIVE PROFILES WITH ONES THAT WERE BURRED AND SANDED SO THEY HAD A SMOOTH EDGE AND DID NOT CAUSE ANY MORE CUTS. THE MANUFACTURER OF THE U PROFILES DETECTED THAT THE PROBLEM CAME FROM A NON SUFFICIENTLY TRAINED EMPLOYEE, HAS SINCE INCREASED HIS QUALITY AND TRAINING PROCESS AND ADAPTED A LASER PROCESSED PRODUCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UT69 TABLE MOUNTED RADIATION SHIELDS UT69 KPY MAVIG GMBH UT6902-US

Patients

Seq Age Sex Outcome Treatment
1 Other