BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE
Report
- Report Number
- 1024879-2023-00901
- Event Type
- Malfunction
- Date Received
- December 14, 2023
- Date of Event
- November 23, 2023
- Report Date
- January 9, 2024
- Manufacturer
- BECTON, DICKINSON & CO., (BD)
- Product Code
- FMI
- UDI-DI
- 50382903686075
- PMA / PMN Number
- K982541
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6: INVESTIGATION SUMMARY: MATERIAL #: 368607. LOT/BATCH #: 3193560. BD HAD NOT RECEIVED SAMPLES, BUT 3 PHOTOS WERE PROVIDED FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR DEFECTIVE LOCKING MECHANISM WAS OBSERVED. ADDITIONALLY, 20 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND FUNCTIONAL TESTING, EACH HAVING THEIR ECLIPSE SHIELD ACTIVATED, AND THE ISSUE OF DEFECTIVE LOCKING MECHANISM WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE DEFECTIVE LOCKING MECHANISM. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD BREAKS OFF THE HINGE AND THE USER CAN'T COMPLETELY SLIDE IT OFF. THIS HAS OCCURRED WITH AN UNSPECIFIED NUMBER OF DEVICES. NO PATIENT IMPACT REPORTED.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD BREAKS OFF THE HINGE AND THE USER CAN'T COMPLETELY SLIDE IT OFF. THIS HAS OCCURRED WITH AN UNSPECIFIED NUMBER OF DEVICES. NO PATIENT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 327608 | BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BECTON, DICKINSON & CO., (BD) | 3193560 | 50382903686075 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |