FDA Adverse Event Malfunction Summary report: N

MAVERICK2 BALLOON CATHETER

MDR report key: 1832820 · Received September 13, 2010

Report

Report Number
2134265-2010-03995
Event Type
Malfunction
Date Received
September 13, 2010
Date of Event
August 23, 2010
Report Date
August 23, 2010
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS THE DEVICE HAS NOT BEEN RETURNED, A TECHNICAL ANALYSIS COULD NOT BE PERFORMED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE, A BALLOON RUPTURE OCCURRED. VASCULAR ACCESS WAS OBTAINED THROUGH THE RADIAL ARTERY. THE TARGET LESION WAS LOCATED IN A SEVERELY CALCIFIED DISTAL LEFT CIRCUMFLEX (CX) ARTERY. THE PHYSICIAN ADVANCED THE 2.5MM X 15MM MAVERICK 2 BALLOON CATHETER. ON THE FIRST INFLATION THE BALLOON WAS PARTIALLY INFLATED TO 12ATMS AND RUPTURED. THE DEVICE WAS REMOVED FROM THE PATIENT INTACT AND THE PROCEDURE WAS COMPLETED WITH A NON-BSC BALLOON. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAVERICK2 BALLOON CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - GALWAY H7493892815250 0013396617

Patients

Seq Age Sex Outcome Treatment
1