FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP

MDR report key: 1832739 · Received September 13, 2010

Report

Report Number
6000001-2007-01052
Event Type
Malfunction
Date Received
September 13, 2010
Date of Event
December 16, 2006
Report Date
December 16, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 10, 2007. EVALUATION SUMMARY: THE INFUSION PUMP WAS TESTED FOR FLOW ACCURACY AT 100 ML/HOUR, THE INFUSION PUMP FAILED THE ACCURACY TEST. THE SPECIFICATION OF THIS DEVICE IS +/-5%. EVALUATION WAS COMPLETED AND THE CONDITION OF FAILED ACCURACY IN CHANNEL B, DUE TO A DAMAGED PUMP HEAD MODULE (PHM), WAS CONFIRMED. THE PHM CHANNEL B WAS REPLACED AND THE BAXTER TECHNICIAN TESTED THE DEVICE.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE BY BAXTER A FAILED ACCURACY TEST IN CHANNEL B WAS REPORTED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP INUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment