FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP
MDR report key: 1832712
·
Received September 13, 2010
Report
- Report Number
- 6000001-2007-01046
- Event Type
- Malfunction
- Date Received
- September 13, 2010
- Date of Event
- December 15, 2006
- Report Date
- December 15, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON (B)(4) 2007. EVALUATION SUMMARY: THE INFUSION PUMP WAS TESTED FOR FLOW ACCURACY AT 100ML/HR, THE PUMP FAILED THE ACCURACY TEST WITH A FLOW VALUE BELOW THE DESIRED AMOUNT. THE SPECIFICATION LIMIT FOR THIS PUMP IS +/-5%. THE PUMPHEAD WAS REPLACED AND THE DEVICE WAS RETURNED TO THE CUSTOMER FULLY OPERATIONAL.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED FROM CUSTOMER FOR SERVICE. DURING SERVICE BY BAXTER TECHNICIAN, THE DEVICE FAILED ACCURACY TEST. CUSTOMER REPORTED THERE WERE NO PATIENT INJURIES OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |