FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP

MDR report key: 1832712 · Received September 13, 2010

Report

Report Number
6000001-2007-01046
Event Type
Malfunction
Date Received
September 13, 2010
Date of Event
December 15, 2006
Report Date
December 15, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON (B)(4) 2007. EVALUATION SUMMARY: THE INFUSION PUMP WAS TESTED FOR FLOW ACCURACY AT 100ML/HR, THE PUMP FAILED THE ACCURACY TEST WITH A FLOW VALUE BELOW THE DESIRED AMOUNT. THE SPECIFICATION LIMIT FOR THIS PUMP IS +/-5%. THE PUMPHEAD WAS REPLACED AND THE DEVICE WAS RETURNED TO THE CUSTOMER FULLY OPERATIONAL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED FROM CUSTOMER FOR SERVICE. DURING SERVICE BY BAXTER TECHNICIAN, THE DEVICE FAILED ACCURACY TEST. CUSTOMER REPORTED THERE WERE NO PATIENT INJURIES OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1