FDA Adverse Event Malfunction Summary report: N

COLLEAGUE VOLUMETRIC INFUSION PUMP

MDR report key: 1832694 · Received September 13, 2010

Report

Report Number
6000001-2007-00412
Event Type
Malfunction
Date Received
September 13, 2010
Date of Event
December 12, 2006
Report Date
December 12, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).EVALUATION SUMMARY:THE DEVICE EVALUATION WAS COMPLETED AND FAILURE CODE 570:320:844:000 (IN EVENT HISTORY), FOUND TO BE DUE TO THE DEPLETED (DAMAGED) BATTERY CONDITION, WAS CONFIRMED. SERVICE INSTALLED NEW BATTERIES AND TESTED THE DEVICE. A REVIEW OF THE COMPLAINT HISTORY REVEALED SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT AND THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA, (B)(4).

Description of Event or Problem · 1

THE DEVICE WAS RECEIVED FOR SERVICE. DURING SERVICE TESTING, FAILURE CODE 570:320:844:0000 (DAMAGED BATTERIES) WAS FOUND IN THE EVENT HISTORY. ACCORDING TO THE HOSPITAL REPRESENTATIVE, THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA